Development of a 3D-Printed Chest Phantom with Simulation of Lung Nodules for Studying Ultra-Low-Dose Computed Tomography Protocols

This study aimed to 3D print a patient-specific chest phantom simulating multiple lung nodules to optimise low-dose Computed Tomography (CT) protocols for lung cancer screening. The chest phantom, which was developed from a single patient’s chest CT images, was fabricated using a variety of materials, including polylactic acid (PLA), Glow-PLA, acrylonitrile butadiene styrene (ABS), and polyurethane resin. The phantom was scanned under different low-dose (LDCT) and ultra-low-dose CT (ULDCT) protocols by varying the kilovoltage peak (kVp) and milliampere-seconds (mAs). Subjective image quality of each scan (656 images) was evaluated by three radiologists using a five-point Likert scale, while objective image quality was assessed using signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Anatomical conformance was assessed by comparing tissue diameters of the phantom and patient scans using Bland–Altman analysis. The phantom’s lung tissue, lung nodules, and diaphragm demonstrated radiation attenuation comparable to patient tissue, as measured in Hounsfield Units (HU). However, significant variations in HU were observed for the skin, subcutaneous fat, muscle, bone, heart, lung vessels, and blood vessels compared to patient tissues, with values ranging from 93.9 HU to −196 HU (p < 0.05). Both SNR and CNR decreased as the effective dose was reduced, with a strong positive linear correlation (r = 0.927 and r = 0.931, respectively, p < 0.001, Jamovi, version 2.3.28). The median subjective image quality score from radiologists was 4, indicating good diagnostic confidence across all CT protocols (κ = −0.398, 95% CI [−0.644 to −0.152], p < 0.002, SPSS Statistics, version 30). An optimal protocol of 80 kVp and 30 mAs was identified for lung nodule detection, delivering a dose of only 0.23 mSv, which represents a 96% reduction compared to standard CT protocols. The measurement error between patient and phantom scans was −0.03 ± 0.14 cm. These findings highlight the potential for significant dose reductions in lung cancer screening programs. Further studies are recommended to improve the phantom by selecting more tissue-equivalent materials.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2024-12-31
DOI
https://doi.org/10.3390/app15010309
Citations
3
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
1.82
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a 3D-Printed Chest Phantom with Simulation of Lung Nodules for Studying Ultra-Low-Dose Computed Tomography Protocols

Chai Hong Yeong, Steven Tran, Jenna Silberstein, Zhonghua Sun et al.
3 citations
Applied Sciences
Advanced Radiotherapy Techniques
1.82
article

Development of a 3D-Printed Chest Phantom with Simulation of Lung Nodules for Studying Ultra-Low-Dose Computed Tomography Protocols

Chai Hong Yeong, Steven Tran, Jenna Silberstein, Zhonghua Sun, Yin How Wong
article en
3 citations

Abstract

This study aimed to 3D print a patient-specific chest phantom simulating multiple lung nodules to optimise low-dose Computed Tomography (CT) protocols for lung cancer screening. The chest phantom, which was developed from a single patient’s chest CT images, was fabricated using a variety of materials, including polylactic acid (PLA), Glow-PLA, acrylonitrile butadiene styrene (ABS), and polyurethane resin. The phantom was scanned under different low-dose (LDCT) and ultra-low-dose CT (ULDCT) protocols by varying the kilovoltage peak (kVp) and milliampere-seconds (mAs). Subjective image quality of each scan (656 images) was evaluated by three radiologists using a five-point Likert scale, while objective image quality was assessed using signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Anatomical conformance was assessed by comparing tissue diameters of the phantom and patient scans using Bland–Altman analysis. The phantom’s lung tissue, lung nodules, and diaphragm demonstrated radiation attenuation comparable to patient tissue, as measured in Hounsfield Units (HU). However, significant variations in HU were observed for the skin, subcutaneous fat, muscle, bone, heart, lung vessels, and blood vessels compared to patient tissues, with values ranging from 93.9 HU to −196 HU (p < 0.05). Both SNR and CNR decreased as the effective dose was reduced, with a strong positive linear correlation (r = 0.927 and r = 0.931, respectively, p < 0.001, Jamovi, version 2.3.28). The median subjective image quality score from radiologists was 4, indicating good diagnostic confidence across all CT protocols (κ = −0.398, 95% CI [−0.644 to −0.152], p < 0.002, SPSS Statistics, version 30). An optimal protocol of 80 kVp and 30 mAs was identified for lung nodule detection, delivering a dose of only 0.23 mSv, which represents a 96% reduction compared to standard CT protocols. The measurement error between patient and phantom scans was −0.03 ± 0.14 cm. These findings highlight the potential for significant dose reductions in lung cancer screening programs. Further studies are recommended to improve the phantom by selecting more tissue-equivalent materials.

Applied SciencesVol. 15(1)
Curtin University (AU), Taylor's University (MY)
Good health and well-being
Openalex Percentile: Top 14%
Advanced Radiotherapy Techniques
1.82
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.