Dose Verification for Linac-Based Stereotactic Radiosurgery planned at Different Prescription Isodose Levels Using Delta4 Phantom+

Background: Linear accelerator (LINAC)-based stereotactic radiosurgery (SRS) plans and their treatment are complex techniques that require a comprehensive quality assurance program before they are clinically implemented. To cope with this intricacy, clinics must comprehensively validate treatment plans to deliver precise doses and assure patients. The study aimed to verify the treatment planning dose to the dose delivered at the LINAC during the SRS treatment planned at different prescription isodoses with the new wireless Delta4 Phantom+.Materials and Methods: Clinically accepted volumetric modulated arc therapy (VMAT) SRS plans made with the Stereotactic End-to-End Verification (STEEV) anthropomorphic phantom were created with six different prescription isodose level using 6 MV flattening filter free (FFF) beam. All these VMAT SRS plans were replicated on the Delta4phantom+ and delivered with Varian Truebeam LINAC. The planned and delivered dose showed excellent correlation, and this was evaluated using distance to agreement (2 mm), dose deviation (2%), and gamma-index passing rate.Results: The results showed that the calculated treatment planning system (TPS) dose and the measurement with the delta4 plus phantom were in excellent accord. The minimum gamma pass rate was 99.6% and the maximum 100%. The gamma passing rate above 95% for all plans and dose goals were achieved. Conclusion: The verification with the Delta4 Phantom+ measurement depicted an excellent correlation with the dose of the SRS treatment plans for the different prescription isodose levels. The wireless Delta 4Phantom+ device is precise and consistent. It is a quickly set-up device, suitable for SRS treatment verification and allows for real-time measurement. However, we do recommend a stricter passing rate for VMAT SRS Plans.

Authors

Institutions

Publication Details

Journal
Global Clinical Engineering Journal
Published
2024-12-30
DOI
https://doi.org/10.31354/globalce.v6i4.198
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dose Verification for Linac-Based Stereotactic Radiosurgery planned at Different Prescription Isodose Levels Using Delta4 Phantom+

Samuel Nii Adu Tagoe, Emmanuel Fiagbedzi, Francis Hasford
Global Clinical Engineering Journal
Advanced Radiotherapy Techniques
article

Dose Verification for Linac-Based Stereotactic Radiosurgery planned at Different Prescription Isodose Levels Using Delta4 Phantom+

Samuel Nii Adu Tagoe, Emmanuel Fiagbedzi, Francis Hasford
article en

Abstract

Background: Linear accelerator (LINAC)-based stereotactic radiosurgery (SRS) plans and their treatment are complex techniques that require a comprehensive quality assurance program before they are clinically implemented. To cope with this intricacy, clinics must comprehensively validate treatment plans to deliver precise doses and assure patients. The study aimed to verify the treatment planning dose to the dose delivered at the LINAC during the SRS treatment planned at different prescription isodoses with the new wireless Delta4 Phantom+.Materials and Methods: Clinically accepted volumetric modulated arc therapy (VMAT) SRS plans made with the Stereotactic End-to-End Verification (STEEV) anthropomorphic phantom were created with six different prescription isodose level using 6 MV flattening filter free (FFF) beam. All these VMAT SRS plans were replicated on the Delta4phantom+ and delivered with Varian Truebeam LINAC. The planned and delivered dose showed excellent correlation, and this was evaluated using distance to agreement (2 mm), dose deviation (2%), and gamma-index passing rate.Results: The results showed that the calculated treatment planning system (TPS) dose and the measurement with the delta4 plus phantom were in excellent accord. The minimum gamma pass rate was 99.6% and the maximum 100%. The gamma passing rate above 95% for all plans and dose goals were achieved. Conclusion: The verification with the Delta4 Phantom+ measurement depicted an excellent correlation with the dose of the SRS treatment plans for the different prescription isodose levels. The wireless Delta 4Phantom+ device is precise and consistent. It is a quickly set-up device, suitable for SRS treatment verification and allows for real-time measurement. However, we do recommend a stricter passing rate for VMAT SRS Plans.

Global Clinical Engineering JournalVol. 6(4)
University of Ghana (GH)
Openalex Percentile: Top 66%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Dose Verification for Linac-Based Stereotactic Radiosurgery planned at Different Prescription Isodose Levels Using Delta4 Phantom+ — Samuel Nii Adu Tagoe, Emmanuel Fiagbedzi, et al. · Global Clinical Engineering Journal (2024) | TGRS Research Map | TGRS