Preliminary Investigation of Electromagnetic Compatibility Analysis for Magnetic Resonance Human Body Communication

This paper presents a preliminary exploration of the electromagnetic distribution and electromagnetic compatibility issues related to Magnetic Resonance Human Body Communication (MR-HBC) technology. Through the design of magnetic resonant coils and finite element simulation analysis, the distribution of magnetic field strength and current density within the human body was evaluated. The results indicate that both the magnetic field strength and current density decrease logarithmically as the depth within the body increases (from 10 mm to 150 mm). Notably, high coupling strength remains relatively stable within a depth of 50 mm. Further analysis confirms that the induced currents in MR-HBC are dominated by variations in the magnetic field, exhibiting the volume conductor characteristics of human tissue under specific conditions. MR-HBC technology demonstrates excellent electromagnetic biocompatibility, with current density at operational frequencies well below safety limits, thereby providing a safe foundation and theoretical basis for the application of implanted and wearable medical devices.

Authors

Institutions

Publication Details

Published
2024-11-04
DOI
https://doi.org/10.1109/csrswtc64338.2024.10811652
Primary Topic
Wireless Body Area Networks
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Preliminary Investigation of Electromagnetic Compatibility Analysis for Magnetic Resonance Human Body Communication

Jiamei Wang, Yueming Gao, Mang I Vai, Sio Hang Pun et al.
2 links
Wireless Body Area Networks
article

Preliminary Investigation of Electromagnetic Compatibility Analysis for Magnetic Resonance Human Body Communication

Jiamei Wang, Yueming Gao, Mang I Vai, Sio Hang Pun, Han Wang, Shujuan Jiang, Dongming Li
article en

Abstract

This paper presents a preliminary exploration of the electromagnetic distribution and electromagnetic compatibility issues related to Magnetic Resonance Human Body Communication (MR-HBC) technology. Through the design of magnetic resonant coils and finite element simulation analysis, the distribution of magnetic field strength and current density within the human body was evaluated. The results indicate that both the magnetic field strength and current density decrease logarithmically as the depth within the body increases (from 10 mm to 150 mm). Notably, high coupling strength remains relatively stable within a depth of 50 mm. Further analysis confirms that the induced currents in MR-HBC are dominated by variations in the magnetic field, exhibiting the volume conductor characteristics of human tissue under specific conditions. MR-HBC technology demonstrates excellent electromagnetic biocompatibility, with current density at operational frequencies well below safety limits, thereby providing a safe foundation and theoretical basis for the application of implanted and wearable medical devices.

University of Macau (MO), Shanghai Public Security Bureau (CN), Fuzhou University (CN)
National Natural Science Foundation of China
Openalex Percentile: Top 77%
Wireless Body Area Networks
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.