Comparing dislocation force between a flanged haptics IOL and a harpoon haptics IOL

To compare two different secondary IOL fixation techniques, either flanged or hooked, regarding the least required force to dislocate the haptic in human corneoscleral donor tissue (CST). Experimental laboratory investigation. The least required dislocation force (LRDF) of two different fixation techniques, namely the flanged haptics (FH, as described by Yamane) and the harpoon haptic technique (HH, as described by Carlevale) were investigated using 20 three-piece IOLs (KOWA PU6AS) and 20 single-piece IOLs (SOLEKO CARLEVALE) fixated to human scleral tissue. The main outcome, differences in LRDF of the investigated techniques, was measured with a tensiometer. The dislocation force needed to dislocate the flanged haptics was significantly higher (p < 0.001) in the flanged 3-piece IOL (0.93 ± 0.43 N) than the specialized, harpoon haptics single-piece IOL (0.45 ± 0.18 N). During externalization, breakage occurred in three harpoon haptics. However, no breakage was observed in either haptics during dislocation. The flanged haptic technique proved to be the stronger form of secondary IOL fixation regarding dislocation force in this in vitro study. The harpoon haptics fixation technique showed significantly less resistance to axial traction and a susceptibility to breakage during externalization.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2024-12-30
DOI
https://doi.org/10.1038/s41598-024-83774-w
Citations
1
Primary Topic
Intraocular Surgery and Lenses
Type
article
Field-Weighted Citation Impact
0.54

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparing dislocation force between a flanged haptics IOL and a harpoon haptics IOL

Andreas Schlatter, Martin Kronschläger, Oliver Findl, Johannes Zeilinger et al.
1 citations
Scientific Reports
Intraocular Surgery and Lenses
0.54
article

Comparing dislocation force between a flanged haptics IOL and a harpoon haptics IOL

Andreas Schlatter, Martin Kronschläger, Oliver Findl, Johannes Zeilinger, Natascha Bayer
article en
1 citations

Abstract

To compare two different secondary IOL fixation techniques, either flanged or hooked, regarding the least required force to dislocate the haptic in human corneoscleral donor tissue (CST). Experimental laboratory investigation. The least required dislocation force (LRDF) of two different fixation techniques, namely the flanged haptics (FH, as described by Yamane) and the harpoon haptic technique (HH, as described by Carlevale) were investigated using 20 three-piece IOLs (KOWA PU6AS) and 20 single-piece IOLs (SOLEKO CARLEVALE) fixated to human scleral tissue. The main outcome, differences in LRDF of the investigated techniques, was measured with a tensiometer. The dislocation force needed to dislocate the flanged haptics was significantly higher (p < 0.001) in the flanged 3-piece IOL (0.93 ± 0.43 N) than the specialized, harpoon haptics single-piece IOL (0.45 ± 0.18 N). During externalization, breakage occurred in three harpoon haptics. However, no breakage was observed in either haptics during dislocation. The flanged haptic technique proved to be the stronger form of secondary IOL fixation regarding dislocation force in this in vitro study. The harpoon haptics fixation technique showed significantly less resistance to axial traction and a susceptibility to breakage during externalization.

Scientific ReportsVol. 14(1)
Hanusch Hospital (AT), Vienna Institute for Research in Ocular Surgery (AT)
Universität Wien, Medizinische Universität Wien
Life below water
Openalex Percentile: Top 28%
Intraocular Surgery and Lenses
0.54
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Comparing dislocation force between a flanged haptics IOL and a harpoon haptics IOL — Andreas Schlatter, Martin Kronschläger, et al. · Scientific Reports (2024) | TGRS Research Map | TGRS