Evaluation of Disinfection Efficacy in Flexible Laryngoscopy According to Cidex Ortho-Phthalaldehyde Immersion Time: A Multicenter Study

Background and Objectives Flexible laryngoscopes are indispensable tools in otolaryngology, but their frequent use makes them vulnerable to contamination, thus posing a risk of cross-infection. Unlike gastrointestinal endoscopes, flexible laryngoscopes currently lack standardized disinfection protocols. This study evaluates the efficacy of Cidex OPA (0.55% ortho-phthalaldehyde) to establish an effective, practical disinfection protocol for flexible laryngoscopes.Materials and Method This multicenter study involved the use of flexible laryngoscopes in otolaryngology outpatient clinics across five university hospitals. Laryngoscopes were immersed in Cidex OPA for 1, 5, or 12 minutes, with an additional group treated using Tristel wipes and foam after a 12-minute immersion. Swab samples were collected from the distal 15 cm of each laryngoscope following disinfection and cultured on blood agar plates under aerobic conditions at 35°C–37°C with 5% CO<sub>2</sub> for 72 hours. Positive controls included laryngoscopes directly contaminated with saliva or laryngeal secretions.Results Six out of ten positive control samples demonstrated bacterial growth. However, no bacterial growth was observed in any sample from the Cidex OPA immersion groups (1, 5, or 12 minutes), including the group treated with Tristel. These findings indicate that even a 1-minute immersion in Cidex OPA effectively eliminates bacterial contamination.Conclusion This study provides evidence supporting an efficient disinfection method that can enhance infection control and streamline clinical workflow. Further research with a larger sample size and varied disinfection techniques is needed to establish comprehensive disinfection guidelines for flexible laryngoscopes.

Authors

Publication Details

Journal
Journal of The Korean Society of Laryngology Phoniatrics and Logopedics
Published
2024-12-30
DOI
https://doi.org/10.22469/jkslp.2024.35.3.89
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Disinfection Efficacy in Flexible Laryngoscopy According to Cidex Ortho-Phthalaldehyde Immersion Time: A Multicenter Study

Young Chan Lee, Nayeon Choi, K. Cho, Sang Hyuk Lee et al.
Journal of The Korean Society of Laryngology Phoniatrics and Logopedics
Medical Device Sterilization and Disinfection
article

Evaluation of Disinfection Efficacy in Flexible Laryngoscopy According to Cidex Ortho-Phthalaldehyde Immersion Time: A Multicenter Study

Young Chan Lee, Nayeon Choi, K. Cho, Sang Hyuk Lee, Heejin Kim, Sei Young Lee, Hae Sang Park
article en

Abstract

Background and Objectives Flexible laryngoscopes are indispensable tools in otolaryngology, but their frequent use makes them vulnerable to contamination, thus posing a risk of cross-infection. Unlike gastrointestinal endoscopes, flexible laryngoscopes currently lack standardized disinfection protocols. This study evaluates the efficacy of Cidex OPA (0.55% ortho-phthalaldehyde) to establish an effective, practical disinfection protocol for flexible laryngoscopes.Materials and Method This multicenter study involved the use of flexible laryngoscopes in otolaryngology outpatient clinics across five university hospitals. Laryngoscopes were immersed in Cidex OPA for 1, 5, or 12 minutes, with an additional group treated using Tristel wipes and foam after a 12-minute immersion. Swab samples were collected from the distal 15 cm of each laryngoscope following disinfection and cultured on blood agar plates under aerobic conditions at 35°C–37°C with 5% CO<sub>2</sub> for 72 hours. Positive controls included laryngoscopes directly contaminated with saliva or laryngeal secretions.Results Six out of ten positive control samples demonstrated bacterial growth. However, no bacterial growth was observed in any sample from the Cidex OPA immersion groups (1, 5, or 12 minutes), including the group treated with Tristel. These findings indicate that even a 1-minute immersion in Cidex OPA effectively eliminates bacterial contamination.Conclusion This study provides evidence supporting an efficient disinfection method that can enhance infection control and streamline clinical workflow. Further research with a larger sample size and varied disinfection techniques is needed to establish comprehensive disinfection guidelines for flexible laryngoscopes.

Journal of The Korean Society of Laryngology Phoniatrics and LogopedicsVol. 35(3)
Openalex Percentile: Top 76%
Medical Device Sterilization and Disinfection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.