Verification of automated review, release and reporting of results with assessment of the risk of harm for patients: the procedure algorithm proposal for clinical laboratories
The autoverified and non-autoverified results showed high agreement with the expert opinions, with minimal disagreement (0.02 % and 0.83 %, respectively). The risk analysis showed that autoverification did not pose a significant risk to patient safety. This study, the first of its kind, provides step-by-step recommendations for implementing autoverification in laboratories.
Authors
- Mario Štefanović (ORCID: https://orcid.org/0000-0002-3965-3231)
- Marijana Miler (ORCID: https://orcid.org/0000-0002-0569-924X)
- Alen Vrtarić (ORCID: https://orcid.org/0000-0002-4104-1402)
- Lara Milevoj Kopčinović (ORCID: https://orcid.org/0000-0003-0360-9148)
- Nora Nikolac (ORCID: https://orcid.org/0000-0001-6831-6309)
- Adriana Unić
- M Bozović
- Ines Vukasović (ORCID: https://orcid.org/0000-0002-7234-3251)
- Gordan Šimić
- Anita Radman
Institutions
- University Hospital Centre Zagreb (HR)
- University of Zagreb (HR)
- Catholic University of Croatia (HR)
Publication Details
- Journal
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Published
- 2024-12-29
- DOI
- https://doi.org/10.1515/cclm-2024-1164
- Citations
- 1
- Primary Topic
- Clinical Laboratory Practices and Quality Control
- Type
- article
- Field-Weighted Citation Impact
- 0.33