Analysis of the interpolation parameters to estimate the errors at specific force values in continuous calibration

The Euramet project 18SIB08 “ComTraForce” developed research to improve the continuous calibration procedures intended to be added to the under review standards that traditionally deals with static force calibration, such as ISO 376 and ISO 7500-1. Among these procedures is the description of a method for the adequate synchronous correlation between the force signals of the reference and the device under test to estimate the indication error for a specific force target value. This method describes the necessity to carry out a double interpolation sequence, using linear least squares adjustments. This paper presents a study on the impact of selecting different ranges of points on the estimation of the interpolation coefficients and the consequent impact on the calculation of the indication errors. The statistical parameters of these curves are also discussed. Experimental data was obtained from measurement series with different force rates carried out in an assembly with a testing machine and a reference sensor.

Authors

Institutions

Publication Details

Journal
Measurement Sensors
Published
2024-12-27
DOI
https://doi.org/10.1016/j.measen.2024.101376
Primary Topic
Advanced Measurement and Metrology Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analysis of the interpolation parameters to estimate the errors at specific force values in continuous calibration

Rafael Soares de Oliveira, Falk Tegtmeier, Andy Knott, Jonas Sander et al.
Measurement Sensors
Advanced Measurement and Metrology Techniques
article

Analysis of the interpolation parameters to estimate the errors at specific force values in continuous calibration

Rafael Soares de Oliveira, Falk Tegtmeier, Andy Knott, Jonas Sander, Wellington Fernandes
article en

Abstract

The Euramet project 18SIB08 “ComTraForce” developed research to improve the continuous calibration procedures intended to be added to the under review standards that traditionally deals with static force calibration, such as ISO 376 and ISO 7500-1. Among these procedures is the description of a method for the adequate synchronous correlation between the force signals of the reference and the device under test to estimate the indication error for a specific force target value. This method describes the necessity to carry out a double interpolation sequence, using linear least squares adjustments. This paper presents a study on the impact of selecting different ranges of points on the estimation of the interpolation coefficients and the consequent impact on the calculation of the indication errors. The statistical parameters of these curves are also discussed. Experimental data was obtained from measurement series with different force rates carried out in an assembly with a testing machine and a reference sensor.

Measurement SensorsVol. 38
Physikalisch-Technische Bundesanstalt (DE), National Physical Laboratory (GB), Instituto Nacional de Metrologia, Qualidade e Tecnologia (BR), Instituto Nacional de Tecnologia (BR)
Openalex Percentile: Top 75%
Advanced Measurement and Metrology Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Analysis of the interpolation parameters to estimate the errors at specific force values in continuous calibration — Rafael Soares de Oliveira, Falk Tegtmeier, et al. · Measurement Sensors (2024) | TGRS Research Map | TGRS