Multi-Pass Arrival Time Correction in Cyclic Ion Mobility Mass Spectrometry for Imaging and Shotgun Lipidomics

Direct-infusion mass spectrometry (DI-MS) and mass spectrometry imaging (MSI) are powerful techniques for lipidomics research. However, annotating isomeric and isobaric lipids with these methods is challenging due to the absence of chromatographic separation. Recently, cyclic ion mobility mass spectrometry (cIM-MS) has been proposed to overcome this limitation. However, fluctuations in room conditions can affect ion mobility multipass arrival times, potentially reducing annotation confidence. In this study, we developed a multipass arrival time correction method that proved effective across various dates, room temperatures, ion mobility settings, and laboratories using mixtures of reference standards. We observed slight variations in the linear correction lines between lipid and nonlipid molecules, underscoring the importance of choosing appropriate reference molecules. Based on these results, we demonstrated that an accurate multipass arrival time database can be constructed from corrected <i>t</i> <sub>0</sub> and <i>t</i> <sub>p</sub> for interlaboratory use and can effectively identify isomeric lipids in MSI using only a single measurement. This approach significantly simplifies the identification process compared to determining multipass collision cross-section, which requires multiple measurements that are both sample- and time-intensive for MSI. Additionally, we validated our multipass drift time correction method in shotgun lipidomics analyses of human and mouse serum samples and observed no matrix effect for the analysis. Despite variations in dates, room temperatures, instruments, and ion mobility settings, our approach reduced the mean drift time differences from over 2% to below 0.2%.

Authors

Institutions

Publication Details

Journal
ACS Measurement Science Au
Published
2024-12-27
DOI
https://doi.org/10.1021/acsmeasuresciau.4c00077
Citations
5
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.20

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Multi-Pass Arrival Time Correction in Cyclic Ion Mobility Mass Spectrometry for Imaging and Shotgun Lipidomics

Yongyut Sirivatanauksorn, Pattipong Wisanpitayakorn, Sakda Khoomrung, Narumol Jariyasopit et al.
5 citations
ACS Measurement Science Au
Mass Spectrometry Techniques and Applications
1.20
article

Multi-Pass Arrival Time Correction in Cyclic Ion Mobility Mass Spectrometry for Imaging and Shotgun Lipidomics

Yongyut Sirivatanauksorn, Pattipong Wisanpitayakorn, Sakda Khoomrung, Narumol Jariyasopit, Martin E. Palmer, Kassaporn Duangkumpha, Jun Xian Goh
article en
5 citations

Abstract

Direct-infusion mass spectrometry (DI-MS) and mass spectrometry imaging (MSI) are powerful techniques for lipidomics research. However, annotating isomeric and isobaric lipids with these methods is challenging due to the absence of chromatographic separation. Recently, cyclic ion mobility mass spectrometry (cIM-MS) has been proposed to overcome this limitation. However, fluctuations in room conditions can affect ion mobility multipass arrival times, potentially reducing annotation confidence. In this study, we developed a multipass arrival time correction method that proved effective across various dates, room temperatures, ion mobility settings, and laboratories using mixtures of reference standards. We observed slight variations in the linear correction lines between lipid and nonlipid molecules, underscoring the importance of choosing appropriate reference molecules. Based on these results, we demonstrated that an accurate multipass arrival time database can be constructed from corrected <i>t</i> <sub>0</sub> and <i>t</i> <sub>p</sub> for interlaboratory use and can effectively identify isomeric lipids in MSI using only a single measurement. This approach significantly simplifies the identification process compared to determining multipass collision cross-section, which requires multiple measurements that are both sample- and time-intensive for MSI. Additionally, we validated our multipass drift time correction method in shotgun lipidomics analyses of human and mouse serum samples and observed no matrix effect for the analysis. Despite variations in dates, room temperatures, instruments, and ion mobility settings, our approach reduced the mean drift time differences from over 2% to below 0.2%.

ACS Measurement Science AuVol. 5(1)
Siriraj Hospital (TH), Waters (United States) (US), Mahidol University (TH), Waters (United Kingdom) (GB)
Mahidol University, Ministry of Higher Education, Science, Research and Innovation, Thailand, Thailand Science Research and Innovation, Faculty of Medicine Siriraj Hospital, Mahidol University, Center of Excellence for Innovation in Chemistry
Affordable and clean energy
Openalex Percentile: Top 22%
Mass Spectrometry Techniques and Applications
1.20
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.