THE BIOMECHANICAL STABILITY OF A NOVEL BIOABSORBABLE MAGNESIUM ALLOY (ZX00:Mg–Zn–Ca) BONE ANCHOR FOR ROTATOR CUFF REPAIRS: A HUMAN EX VIVO STUDY

Introduction: The purpose of this study was to ascertain the primary ex vivo biomechanical stability of a novel bioabsorbable magnesium alloy (ZX00: Mg–Zn–Ca) bone anchor in human cadaveric proximal humeri, indicated in the reconstruction of the rotator cuff. Methods: Twenty human Thiel-embalmed cadaveric humeri were prepared and freed from soft tissue. One [Formula: see text]5.7 × 20.5-mm ZX00 anchor and one [Formula: see text]5.5 mm × 16.3 mm Arthrex Titanium FT Corkscrew (ATC) control anchor were inserted into the footprint of the supraspinatus tendon, 15 mm apart. The humeri were mounted onto a material testing machine and following a 40 N preload, cyclic loading was performed over 400 cycles. If the construct remained intact, ultimate load to failure (ULTF) was measured using an increasing axial load of 1 mm/s, ULTF and mode of failure were recorded. Results: No difference was found in the ability to withstand cyclic loading, mode or load-to-failure strengths between ZX00 and control anchors. The maximum tractional force loaded for the ZX00 anchors had a median of 257.4 N (range 165.3–328.2). The corresponding value for the ATC anchors averaged 239.9 N (range 118.9–306.7). Conclusion: ZX00 alloy anchors appear to provide adequate initial biomechanical stability when compared to an industry-standard control in a cadaveric rotator cuff repair model.

Authors

Institutions

Publication Details

Journal
Journal of Musculoskeletal Research
Published
2024-12-27
DOI
https://doi.org/10.1142/s0218957724500234
Primary Topic
Shoulder and Clavicle Injuries
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

THE BIOMECHANICAL STABILITY OF A NOVEL BIOABSORBABLE MAGNESIUM ALLOY (ZX00:Mg–Zn–Ca) BONE ANCHOR FOR ROTATOR CUFF REPAIRS: A HUMAN EX VIVO STUDY

Niels Hammer, Stefan Schleifenbaum, Stefan Fischerauer, Paul Siegert et al.
Journal of Musculoskeletal Research
Shoulder and Clavicle Injuries
article

THE BIOMECHANICAL STABILITY OF A NOVEL BIOABSORBABLE MAGNESIUM ALLOY (ZX00:Mg–Zn–Ca) BONE ANCHOR FOR ROTATOR CUFF REPAIRS: A HUMAN EX VIVO STUDY

Niels Hammer, Stefan Schleifenbaum, Stefan Fischerauer, Paul Siegert, Annelie‐Martina Weinberg, Christian Wurnig, Rudolf Maximilian Kinsky, Maxim Honea
article en

Abstract

Introduction: The purpose of this study was to ascertain the primary ex vivo biomechanical stability of a novel bioabsorbable magnesium alloy (ZX00: Mg–Zn–Ca) bone anchor in human cadaveric proximal humeri, indicated in the reconstruction of the rotator cuff. Methods: Twenty human Thiel-embalmed cadaveric humeri were prepared and freed from soft tissue. One [Formula: see text]5.7 × 20.5-mm ZX00 anchor and one [Formula: see text]5.5 mm × 16.3 mm Arthrex Titanium FT Corkscrew (ATC) control anchor were inserted into the footprint of the supraspinatus tendon, 15 mm apart. The humeri were mounted onto a material testing machine and following a 40 N preload, cyclic loading was performed over 400 cycles. If the construct remained intact, ultimate load to failure (ULTF) was measured using an increasing axial load of 1 mm/s, ULTF and mode of failure were recorded. Results: No difference was found in the ability to withstand cyclic loading, mode or load-to-failure strengths between ZX00 and control anchors. The maximum tractional force loaded for the ZX00 anchors had a median of 257.4 N (range 165.3–328.2). The corresponding value for the ATC anchors averaged 239.9 N (range 118.9–306.7). Conclusion: ZX00 alloy anchors appear to provide adequate initial biomechanical stability when compared to an industry-standard control in a cadaveric rotator cuff repair model.

Journal of Musculoskeletal ResearchVol. 28(01)
Medical University of Graz (AT), Graz University of Technology (AT), RHI Magnesita (Austria) (AT), Fraunhofer Institute for Machine Tools and Forming Technology (DE), University Hospital Leipzig (DE), Orthopaedic Hospital Speising (AT)
Openalex Percentile: Top 64%
Shoulder and Clavicle Injuries
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.