Development and validation of NEMA NU-2 equivalent sensitivity and scatter phantoms for PET performance characteristic evaluation

Abstract Objective This study aimed to evaluate the physical performance of a Philips Gemini TOF PET/CT system using in-house developed sensitivity and scatter phantoms. The assessment was conducted in accordance with the NEMA NU2-2007 guidelines. Methods We developed in-house NEMA NU-2 Sensitivity (NU-2IndiSen) and Scatter (NU-2IndiSc) phantoms. These phantoms were evaluated through a series of tests: the alpha test, which included dimension comparison and radiation exposure comparison, and the beta test, which involved comparing NEMA NU-2 sensitivity and scatter fractions. In the dimension test, we compared the diameters of the NU-2IndiSen and NU-2IndiSc phantoms with those of the NU-2Sen and NU-2Sc phantoms. The radiation exposure tests involved comparing the surface radiation exposure from the NU-2IndiSen and NU-2IndiSc phantoms with that from the NU-2Sen and NU-2Sc phantoms. The beta tests assessed the sensitivity and scatter fraction of the Gemini TF 64 PET system using the NU-2IndiSen and NU-2IndiSc phantoms and compared these measurements with those obtained using the NU-2Sen and NU-2Sc phantoms. Results The dimension tests revealed no significant differences (p-value > 0.05) between the NU-2IndiSen and NU-2Sen phantoms, as well as between the NU-2IndiSc and NU-2Sc phantoms. Similarly, the radiation exposure tests showed no significant differences (p-value > 0.05) in surface radiation exposure between the NU-2IndiSen and NU-2Sen phantoms, and between the NU-2IndiSc and NU-2Sc phantoms. The average system sensitivity measured using the NU-2IndiSen and NU-2Sen NEMA phantoms were 6724 cps/MBq and 6748 cps/MBq, respectively, both exceeding the vendor's acceptance limit of > 6700 cps/MBq. Similarly, the scatter fraction, noise equivalent count rate (NECR), activity at which peak NECR is achieved, and maximum true count rate performance measured using the NU-2IndiSc and NU-2Sc phantoms were 25.9, 106.4 kcps, 18.5 kBq/ml, and 243.2 kcps, and 27.14, 106.03 kcps, 15.68 kBq/ml, and 263.2 kcps, respectively, all well within the vendor's prescribed limits. Conclusion The results of this study demonstrate that the in-house developed sensitivity and scatter phantoms accurately assess the performance of the Philips Gemini TOF PET/CT system. The obtained values for system sensitivity, scatter fraction, NECR, and true count rate are in excellent agreement with the vendor-specified limits, validating the use of these phantoms for routine quality assurance procedures.

Authors

Institutions

Publication Details

Journal
Health and Technology
Published
2024-12-27
DOI
https://doi.org/10.1007/s12553-024-00930-y
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and validation of NEMA NU-2 equivalent sensitivity and scatter phantoms for PET performance characteristic evaluation

Venkatesh Rangarajan, R. S. Vishwakarma, B Rawat, Viraj Sawant et al.
1 citations
Health and Technology
Medical Imaging Techniques and Applications
0.40
article

Development and validation of NEMA NU-2 equivalent sensitivity and scatter phantoms for PET performance characteristic evaluation

Venkatesh Rangarajan, R. S. Vishwakarma, B Rawat, Viraj Sawant, Ashish Kumar Jha, Sridhar Sahoo, T. Palani Selvam
article en
1 citations

Abstract

Abstract Objective This study aimed to evaluate the physical performance of a Philips Gemini TOF PET/CT system using in-house developed sensitivity and scatter phantoms. The assessment was conducted in accordance with the NEMA NU2-2007 guidelines. Methods We developed in-house NEMA NU-2 Sensitivity (NU-2IndiSen) and Scatter (NU-2IndiSc) phantoms. These phantoms were evaluated through a series of tests: the alpha test, which included dimension comparison and radiation exposure comparison, and the beta test, which involved comparing NEMA NU-2 sensitivity and scatter fractions. In the dimension test, we compared the diameters of the NU-2IndiSen and NU-2IndiSc phantoms with those of the NU-2Sen and NU-2Sc phantoms. The radiation exposure tests involved comparing the surface radiation exposure from the NU-2IndiSen and NU-2IndiSc phantoms with that from the NU-2Sen and NU-2Sc phantoms. The beta tests assessed the sensitivity and scatter fraction of the Gemini TF 64 PET system using the NU-2IndiSen and NU-2IndiSc phantoms and compared these measurements with those obtained using the NU-2Sen and NU-2Sc phantoms. Results The dimension tests revealed no significant differences (p-value > 0.05) between the NU-2IndiSen and NU-2Sen phantoms, as well as between the NU-2IndiSc and NU-2Sc phantoms. Similarly, the radiation exposure tests showed no significant differences (p-value > 0.05) in surface radiation exposure between the NU-2IndiSen and NU-2Sen phantoms, and between the NU-2IndiSc and NU-2Sc phantoms. The average system sensitivity measured using the NU-2IndiSen and NU-2Sen NEMA phantoms were 6724 cps/MBq and 6748 cps/MBq, respectively, both exceeding the vendor's acceptance limit of > 6700 cps/MBq. Similarly, the scatter fraction, noise equivalent count rate (NECR), activity at which peak NECR is achieved, and maximum true count rate performance measured using the NU-2IndiSc and NU-2Sc phantoms were 25.9, 106.4 kcps, 18.5 kBq/ml, and 243.2 kcps, and 27.14, 106.03 kcps, 15.68 kBq/ml, and 263.2 kcps, respectively, all well within the vendor's prescribed limits. Conclusion The results of this study demonstrate that the in-house developed sensitivity and scatter phantoms accurately assess the performance of the Philips Gemini TOF PET/CT system. The obtained values for system sensitivity, scatter fraction, NECR, and true count rate are in excellent agreement with the vendor-specified limits, validating the use of these phantoms for routine quality assurance procedures.

Health and TechnologyVol. 15(2)
Bhabha Atomic Research Centre (IN), Homi Bhabha National Institute (IN), Tata Memorial Hospital (IN), Advanced Centre for Treatment, Research and Education in Cancer (IN)
Openalex Percentile: Top 31%
Medical Imaging Techniques and Applications
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.