Investigation of focus variation microscopy immunity to vibrations

Optical metrology plays a vital role in a wide range of research fields and for inspection in manufacturing industries. At present, the market offers a variety of optical metrology instruments, such as interferometers, confocal microscopes, and focus variation (FV) instruments. Although interferometers have the highest precision among optical metrology instruments, they are very sensitive to vibrations/environmental disturbances. On the other hand, focus variation technology is widely recognised for its robustness to vibrations (compared to interferometers), but so far, there is no study in place investigating the vibration frequency and amplitude limits within which focus variation instruments operate optimally in the presence of vibrations. To our knowledge, this article is the first study that aims to estimate quantitatively the immunity of focus variation instruments to vibration. This paper presents theoretical simulations to investigate how vibrations affect the FV principle of evaluating the best-focused images to calculate surface topography. The simulations were verified with practical results from an experimental FV setup built in the lab, and the two results match very well. Afterwards, vibration experiments were performed using the state-of-the-art focus variation instrument, Alicona InfiniteFocus G5, to measure the surface roughness of the Microusurf 334 comparator from Rubert & Co. LTD under vibrations induced by the P-840.2 piezoelectric actuator from Physik Instrumente (PI). The experiments were performed at different frequencies by incrementally changing the vibration amplitudes, pre-planned as a function of the depth of field (DoF) of each magnification lens (10x to 100x). It is observed that the FV system generates 100 % “bad-data” when the vibration amplitude exceeds three times the DoF of the used objective lens at low frequencies (i.e. as early as 5 Hz). • Simulations and experimental investigations to analyse the impact of vibrations on Focus Variation (FV) Microscope. • FV systems generate “bad-data” when vibration amplitude exceeds three times Depth of Field (DoF) of the objective lens. • Reducing the vertical resolution of the FV system extends the vibration amplitude threshold to six times the DoF.

Authors

Institutions

Publication Details

Journal
Precision Engineering
Published
2024-12-25
DOI
https://doi.org/10.1016/j.precisioneng.2024.12.014
Citations
1
Primary Topic
Image Processing Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.32

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigation of focus variation microscopy immunity to vibrations

Dawei Tang, Hussam Muhamedsalih, Aalim M. Mustafa, Liam Blunt et al.
1 citations
Precision Engineering
Image Processing Techniques and Applications
0.32
article

Investigation of focus variation microscopy immunity to vibrations

Dawei Tang, Hussam Muhamedsalih, Aalim M. Mustafa, Liam Blunt, Prashant Kumar, Jane Jiang
article en
1 citations

Abstract

Optical metrology plays a vital role in a wide range of research fields and for inspection in manufacturing industries. At present, the market offers a variety of optical metrology instruments, such as interferometers, confocal microscopes, and focus variation (FV) instruments. Although interferometers have the highest precision among optical metrology instruments, they are very sensitive to vibrations/environmental disturbances. On the other hand, focus variation technology is widely recognised for its robustness to vibrations (compared to interferometers), but so far, there is no study in place investigating the vibration frequency and amplitude limits within which focus variation instruments operate optimally in the presence of vibrations. To our knowledge, this article is the first study that aims to estimate quantitatively the immunity of focus variation instruments to vibration. This paper presents theoretical simulations to investigate how vibrations affect the FV principle of evaluating the best-focused images to calculate surface topography. The simulations were verified with practical results from an experimental FV setup built in the lab, and the two results match very well. Afterwards, vibration experiments were performed using the state-of-the-art focus variation instrument, Alicona InfiniteFocus G5, to measure the surface roughness of the Microusurf 334 comparator from Rubert & Co. LTD under vibrations induced by the P-840.2 piezoelectric actuator from Physik Instrumente (PI). The experiments were performed at different frequencies by incrementally changing the vibration amplitudes, pre-planned as a function of the depth of field (DoF) of each magnification lens (10x to 100x). It is observed that the FV system generates 100 % “bad-data” when the vibration amplitude exceeds three times the DoF of the used objective lens at low frequencies (i.e. as early as 5 Hz). • Simulations and experimental investigations to analyse the impact of vibrations on Focus Variation (FV) Microscope. • FV systems generate “bad-data” when vibration amplitude exceeds three times Depth of Field (DoF) of the objective lens. • Reducing the vertical resolution of the FV system extends the vibration amplitude threshold to six times the DoF.

Precision EngineeringVol. 93
University of Huddersfield (GB)
Engineering and Physical Sciences Research Council
Openalex Percentile: Top 34%
Image Processing Techniques and Applications
0.32
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.