Sensitivity calibration for mass comparator based on optical force

The laser photon-momentum-based method is more and more used in the optical power measurements. As a revert use, the laser optical power can be used as a traceable source to small mass in micro range using the same photon-momentum-based method to new SI with better measuring accuracy. The sensitivity calibration for mass comparator based on laser optical force was carried out using an optical power range between 20 W and 120 W at a wavelength of 1080 nm. The air density meter is used to sense the stability of the laser power and optical force. Uncertainty of measurement is evaluated. With a coverage factor k = 2, relative expanded uncertainty of reference sensitivity mass standard using photon force of laser is better than 2 %. The uncertainty of the sensitivity calibration is also investigated.

Authors

Institutions

Publication Details

Journal
Measurement Sensors
Published
2024-12-24
DOI
https://doi.org/10.1016/j.measen.2024.101366
Primary Topic
Mechanical and Optical Resonators
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Sensitivity calibration for mass comparator based on optical force

Marian Y. Hu, Hanghang Chen, Jian Wang, Qing Sun et al.
Measurement Sensors
Mechanical and Optical Resonators
article

Sensitivity calibration for mass comparator based on optical force

Marian Y. Hu, Hanghang Chen, Jian Wang, Qing Sun, Changqing Cai, Kai Jiao
article en

Abstract

The laser photon-momentum-based method is more and more used in the optical power measurements. As a revert use, the laser optical power can be used as a traceable source to small mass in micro range using the same photon-momentum-based method to new SI with better measuring accuracy. The sensitivity calibration for mass comparator based on laser optical force was carried out using an optical power range between 20 W and 120 W at a wavelength of 1080 nm. The air density meter is used to sense the stability of the laser power and optical force. Uncertainty of measurement is evaluated. With a coverage factor k = 2, relative expanded uncertainty of reference sensitivity mass standard using photon force of laser is better than 2 %. The uncertainty of the sensitivity calibration is also investigated.

Measurement SensorsVol. 38
National Institute of Metrology (CN)
Affordable and clean energy
Openalex Percentile: Top 78%
Mechanical and Optical Resonators
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Sensitivity calibration for mass comparator based on optical force — Marian Y. Hu, Hanghang Chen, et al. · Measurement Sensors (2024) | TGRS Research Map | TGRS