A new tibial insert design with ball-in-socket medial conformity and posterior cruciate ligament retention has low tibial baseplate migration after unrestricted kinematically aligned total knee arthroplasty: a cohort study using radiostereometric ana
Using the new BS MC insert with PCL retention, the tibial baseplate was stable at the group level at 1 year. Baseplate stability was not strongly related to varus baseplate and limb alignment. Comparable patient-reported outcome scores and maximum flexion/extension at 1 year were shown between the 2 insert designs.
Authors
- Abigail E. Niesen (ORCID: https://orcid.org/0000-0002-7350-7940)
- Maury L. Hull (ORCID: https://orcid.org/0000-0001-8305-4589)
- Stephen Miller Howell (ORCID: https://orcid.org/0000-0001-8184-0324)
- Pranav A. Tirumalai
Institutions
- University of California Davis Medical Center (US)
- University of California, Davis (US)
Publication Details
- Journal
- Acta Orthopaedica
- Published
- 2024-12-23
- DOI
- https://doi.org/10.2340/17453674.2024.42489
- Citations
- 7
- Primary Topic
- Total Knee Arthroplasty Outcomes
- Type
- article
- Field-Weighted Citation Impact
- 4.01