Benchtop centrifugation: An effective method for reducing lipaemia associated interference in grossly lipaemic samples?

BackgroundGrossly lipaemic samples are a significant cause of analytical errors, potentially impacting patient care. The causes of lipaemia are varied and often unavoidable, while methods to reduce lipaemia through gold-standard ultracentrifugation are limited by availability, transportation and cost. Benchtop centrifugation has been proposed as an alternative method to reduce lipaemia.MethodsFifty-three grossly lipaemic serum samples (lipaemia >201.8 mg/dL) that were unsuitable for analysis were selected and centrifugated at 18840 g for different time-periods with lipaemia measured prior to and after centrifugation. Core analytes were measured on serum samples free of lipaemia before and after 30 min of centrifugation at 18840 g to assess the effect of the centrifugation process.ResultsAfter centrifugation for 5 min, 90% of grossly lipaemic samples were either ideal (lipaemia <50 mg/dL) or adequate (lipaemia 50.1-201.7 mg/dL) for analyte testing. All samples were either adequate or ideal for testing following centrifugation for 30 min. Although some analytes showed a statistically significant change in the measured concentration post high-speed centrifugation, none had clinically significant changes according to analyte specific reference change value (RCV) analysis. Aspartate aminotransferase (AST) and creatine kinase (CK) demonstrated the most notable reductions in activity, but these did not exceed their RCV.ConclusionsBenchtop centrifugation shows potential laboratory utility in reducing lipaemia whilst maintaining clinically reliable results, however small sample sizes preclude firm conclusions. Further research is warranted to increase the sample size with finer time-point tuning, sub-group analysis and temperature analysis, due to the potential for sample heat injury, to balance practicality and accuracy.

Authors

Institutions

Publication Details

Journal
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Published
2024-12-23
DOI
https://doi.org/10.1177/00045632241312622
Citations
1
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.33
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Benchtop centrifugation: An effective method for reducing lipaemia associated interference in grossly lipaemic samples?

Kate E. Shipman, Henry Carlton, James Hall
1 citations
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Clinical Laboratory Practices and Quality Control
0.33
article

Benchtop centrifugation: An effective method for reducing lipaemia associated interference in grossly lipaemic samples?

Kate E. Shipman, Henry Carlton, James Hall
article en
1 citations

Abstract

BackgroundGrossly lipaemic samples are a significant cause of analytical errors, potentially impacting patient care. The causes of lipaemia are varied and often unavoidable, while methods to reduce lipaemia through gold-standard ultracentrifugation are limited by availability, transportation and cost. Benchtop centrifugation has been proposed as an alternative method to reduce lipaemia.MethodsFifty-three grossly lipaemic serum samples (lipaemia >201.8 mg/dL) that were unsuitable for analysis were selected and centrifugated at 18840 g for different time-periods with lipaemia measured prior to and after centrifugation. Core analytes were measured on serum samples free of lipaemia before and after 30 min of centrifugation at 18840 g to assess the effect of the centrifugation process.ResultsAfter centrifugation for 5 min, 90% of grossly lipaemic samples were either ideal (lipaemia <50 mg/dL) or adequate (lipaemia 50.1-201.7 mg/dL) for analyte testing. All samples were either adequate or ideal for testing following centrifugation for 30 min. Although some analytes showed a statistically significant change in the measured concentration post high-speed centrifugation, none had clinically significant changes according to analyte specific reference change value (RCV) analysis. Aspartate aminotransferase (AST) and creatine kinase (CK) demonstrated the most notable reductions in activity, but these did not exceed their RCV.ConclusionsBenchtop centrifugation shows potential laboratory utility in reducing lipaemia whilst maintaining clinically reliable results, however small sample sizes preclude firm conclusions. Further research is warranted to increase the sample size with finer time-point tuning, sub-group analysis and temperature analysis, due to the potential for sample heat injury, to balance practicality and accuracy.

Annals of Clinical Biochemistry International Journal of Laboratory MedicineVol. 62(4)
Brighton and Sussex Medical School (GB), University of Sussex (GB), St Richard's Hospital (GB), University Hospitals Sussex NHS Foundation Trust (GB)
Clean water and sanitation
Openalex Percentile: Top 35%
Clinical Laboratory Practices and Quality Control
0.33
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.