A rapid and highly sensitive fluorescence immunochromatographic test strip for pepsin detection in human hypopharyngeal saliva

Pepsin serves as a potential biomarker for laryngopharyngeal reflux disease (LPRD). The reported methods for detecting pepsin still have limitations, such as long reaction times and low sensitivity. Therefore, it is crucial to develop a rapid, simple and sensitive detection method. In this study, a fluorescence immunochromatographic test strip was designed for point-of-care diagnostics of LPRD. To obtain optimal fluorescence immunochromatographic conditions, key experimental factors, including microsphere particle size, binding pH, and resuspension dilution ratio, were systematically evaluated. The optimal fluorescent microsphere size was determined to be 300 nm, the most suitable reaction pH was 8.0, and the ideal resuspension solution to storage solution ratio was 1 : 4. Under these optimal conditions, the sensor had a linear range of 2.5-100.0 ng mL<sup>-1</sup> and a detection limit of 1.9 ng mL<sup>-1</sup>. The relative standard deviation was below 15% for both intra- and inter-batch reproducibility. Under accelerated damage conditions at 37 °C, the test value of the test strip could be kept basically unchanged for 14 days. There was no cross-reactivity of the test strip with pepsinogen I, pepsinogen II and other interferents. In the actual sample detection, the recovery rate was between 85% and 115%. Furthermore, the proposed method was demonstrated to be green and to exhibit strong applicability based on the evaluation using the AGREEprep and the BAGI. In addition, the method has the advantages of high sensitivity, accuracy, greenness and applicability, which can meet the needs of rapid quantitative on-site detection.

Authors

Institutions

Publication Details

Journal
Analytical Methods
Published
2024-12-23
DOI
https://doi.org/10.1039/d4ay01833h
Citations
1
Primary Topic
Gastroesophageal reflux and treatments
Type
article
Field-Weighted Citation Impact
0.46
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A rapid and highly sensitive fluorescence immunochromatographic test strip for pepsin detection in human hypopharyngeal saliva

Juan Huang, Zhenyu Lin, Ting Chen, Jialing Zhang et al.
1 citations
Analytical Methods
Gastroesophageal reflux and treatments
0.46
article

A rapid and highly sensitive fluorescence immunochromatographic test strip for pepsin detection in human hypopharyngeal saliva

Juan Huang, Zhenyu Lin, Ting Chen, Jialing Zhang, Lihong Chen, Shuting Chen, Yaqi Liu
article en
1 citations

Abstract

Pepsin serves as a potential biomarker for laryngopharyngeal reflux disease (LPRD). The reported methods for detecting pepsin still have limitations, such as long reaction times and low sensitivity. Therefore, it is crucial to develop a rapid, simple and sensitive detection method. In this study, a fluorescence immunochromatographic test strip was designed for point-of-care diagnostics of LPRD. To obtain optimal fluorescence immunochromatographic conditions, key experimental factors, including microsphere particle size, binding pH, and resuspension dilution ratio, were systematically evaluated. The optimal fluorescent microsphere size was determined to be 300 nm, the most suitable reaction pH was 8.0, and the ideal resuspension solution to storage solution ratio was 1 : 4. Under these optimal conditions, the sensor had a linear range of 2.5-100.0 ng mL<sup>-1</sup> and a detection limit of 1.9 ng mL<sup>-1</sup>. The relative standard deviation was below 15% for both intra- and inter-batch reproducibility. Under accelerated damage conditions at 37 °C, the test value of the test strip could be kept basically unchanged for 14 days. There was no cross-reactivity of the test strip with pepsinogen I, pepsinogen II and other interferents. In the actual sample detection, the recovery rate was between 85% and 115%. Furthermore, the proposed method was demonstrated to be green and to exhibit strong applicability based on the evaluation using the AGREEprep and the BAGI. In addition, the method has the advantages of high sensitivity, accuracy, greenness and applicability, which can meet the needs of rapid quantitative on-site detection.

Analytical MethodsVol. 17(5)
Fujian Medical University (CN), Fujian Tobacco Industry Limited Liability Company (China) (CN), Fujian Provincial Hospital (CN), Fuzhou University (CN)
Good health and well-being
Openalex Percentile: Top 30%
Gastroesophageal reflux and treatments
0.46
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.