A Low Noise Current Readout with 124 dB Dynamic Range for Bioluminescence Sensing
This paper presents a lower-power photocurrent measurement system for biomolecule quantification. The designed readout comprises a proposed capacitive feedback transimpedance amplifier (C-TIA) in conjunction with an asynchronous flip switch and a self-timed reset network followed by a reconfigurable time-domain quantization scheme. Unlike conventional C-TIAs, this readout provides frequency output without requiring an external clock to perform a periodic reset. Fabricated in a standard 180 nm CMOS process, the readout achieves a current measurement range of 124 dB spanning from $2 \mathrm{pA} \sim 3.5 \mu \mathrm{~A}$, a 98.3% linearity, and a 1 pArms input-referred noise, while consuming only $378 \mu \mathrm{~W}$. In vitro NanoLuc (NLuc) luciferase-based bioluminescence sensing for biomolecule quantification is demonstrated using this chip.
Authors
- Kabir H. Biswas (ORCID: https://orcid.org/0000-0001-9194-4127)
- Amine Bermak (ORCID: https://orcid.org/0000-0003-4984-6093)
- Muhammad Asfandyar Awan (ORCID: https://orcid.org/0000-0003-2464-0523)
- Bo Wang (ORCID: https://orcid.org/0000-0002-9359-4869)
- Samar Shurbaji
Institutions
- Hamad bin Khalifa University (QA)
Publication Details
- Published
- 2024-10-24
- DOI
- https://doi.org/10.1109/biocas61083.2024.10798344
- Citations
- 1
- Primary Topic
- bioluminescence and chemiluminescence research
- Type
- article
- Field-Weighted Citation Impact
- 0.23
Funders
- Qatar National Research Fund