Evaluation of a Continuous Blood Glucose Sensor’s Performance for Hospitalized Patients
Frequent blood glucose monitoring is crucial for managing blood glucose levels in critically ill hospitalized patients experiencing hyperglycemia or hypoglycemia. Existing blood glucose monitoring methods are often cumbersome, painful, and impractical for hourly testing. This work provides a solution for frequent glucose monitoring (less than 1 h per test) for hospitalized patients, ensuring accuracy while minimizing discomfort from blood collection. The glucose sensor demonstrated accuracy within 10% across a range of 0–20 mM over 96 testing cycles. This meets the need for hourly monitoring during an average 2-day hospital stay.
Authors
- Ming‐Yuan Cheng (ORCID: https://orcid.org/0000-0001-6322-1255)
- Ruiqi Lim
- Siti Rafeah Mohamed Rafei
- James Ven Wee Yap
Institutions
- Agency for Science, Technology and Research (SG)
- Institute of Microelectronics (SG)
Publication Details
- Published
- 2024-12-20
- DOI
- https://doi.org/10.3390/engproc2024078005
- Primary Topic
- Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Agency for Science, Technology and Research
- Singapore General Hospital