Rapid Removal of IgG1 Carryover on Protease Column Using Protease-Safe Wash Solutions Delivered with LC Pump for HDX-MS Systems

Sample carryover is a common problem in hydrogen-deuterium exchange mass spectrometry, particularly because immobilized protease columns cannot withstand the high organic solvent concentrations typically used in liquid chromatography-mass spectrometry (LC-MS) for cleaning. Conventional cleaning methods using injections of guanidine HCl still suffer from carryover and may require four blanks after each sample run to fully remove carryover. We have implemented an additional LC pump to deliver customized wash solutions to protease and enzyme columns, and the associated LC capillaries to eliminate carryover. Pump-based washing using the protease-safe wash solutions tested herein was able to fully remove carryover with only one blank run. FOS-choline-12 was found to be the most effective component in wash solutions and even performed well alone at 0.1% volume concentration. Since the protease column washing is performed concurrently during the analytical gradient within a sample run, subsequent blank runs could be reduced from four to one and total run time could be reduced by up to 60%. Savings in total run time could more than double the productivity of data acquisition, which is imperative for pandemic preparedness and for acceleration of biotherapeutics development.

Authors

Institutions

Publication Details

Journal
Journal of the American Society for Mass Spectrometry
Published
2024-12-23
DOI
https://doi.org/10.1021/jasms.4c00411
Citations
2
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.48

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Rapid Removal of IgG1 Carryover on Protease Column Using Protease-Safe Wash Solutions Delivered with LC Pump for HDX-MS Systems

Sandeep Kumar, Kyle W. Anderson
2 citations
Journal of the American Society for Mass Spectrometry
Mass Spectrometry Techniques and Applications
0.48
article

Rapid Removal of IgG1 Carryover on Protease Column Using Protease-Safe Wash Solutions Delivered with LC Pump for HDX-MS Systems

Sandeep Kumar, Kyle W. Anderson
article en
2 citations

Abstract

Sample carryover is a common problem in hydrogen-deuterium exchange mass spectrometry, particularly because immobilized protease columns cannot withstand the high organic solvent concentrations typically used in liquid chromatography-mass spectrometry (LC-MS) for cleaning. Conventional cleaning methods using injections of guanidine HCl still suffer from carryover and may require four blanks after each sample run to fully remove carryover. We have implemented an additional LC pump to deliver customized wash solutions to protease and enzyme columns, and the associated LC capillaries to eliminate carryover. Pump-based washing using the protease-safe wash solutions tested herein was able to fully remove carryover with only one blank run. FOS-choline-12 was found to be the most effective component in wash solutions and even performed well alone at 0.1% volume concentration. Since the protease column washing is performed concurrently during the analytical gradient within a sample run, subsequent blank runs could be reduced from four to one and total run time could be reduced by up to 60%. Savings in total run time could more than double the productivity of data acquisition, which is imperative for pandemic preparedness and for acceleration of biotherapeutics development.

Journal of the American Society for Mass SpectrometryVol. 36(2)
National Institute of Standards and Technology (US), National Institute of Standards (EG), Research Institute for Bioscience and Biotechnology (NP), Material Measurement Laboratory (US), Industrial Measurement Systems (United States) (US), Institute for Bioscience and Biotechnology Research, University of Maryland, College Park (US)
National Institute of Standards and Technology
Openalex Percentile: Top 38%
Mass Spectrometry Techniques and Applications
0.48
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Rapid Removal of IgG1 Carryover on Protease Column Using Protease-Safe Wash Solutions Delivered with LC Pump for HDX-MS Systems — Sandeep Kumar, Kyle W. Anderson · Journal of the American Society for Mass Spectrometry (2024) | TGRS Research Map | TGRS