Development of absolute dosimeter for standardization of low-energy synchrotron radiation and establishing of diagnostic X-ray beam qualities

ABSTRACT Purpose: Low energy X-rays from synchrotron radiation (SR) and X-ray generators are used for diagnostic imaging procedures. The dosimetry of low energy X-rays is challenging. For these types of radiations, the free-air ionization chamber (FAIC) is considered as the primary standard for absolute determination of air-kerma, which is an important quantity for the patient dose during diagnostic radiology investigation. Materials and Methods: To fulfil above requirement, a miniature parallel-plate FAIC (FAIC SR ) and portable parallel-plate FAIC (FAIC DIAG ) were designed and fabricated as the absolute dosimeter for standardization of SR beams and for establishing diagnostic X-rays beam qualities, RQR series respectively. Experiments were performed in SR beams (8-25 keV) as well as in diagnostic X-ray beams of 40, 50, 60, 70, 81, 90, 100 KVp for evaluating the basic characteristics of these chambers. Results: The basic characteristics of these chambers were in conformity with International Atomic Energy Agency (IAEA) TRS-374 and International Electrotechnical Commission (IEC) 61674, 60731. Dosimetric beam quality parameters such as the first half-value layer and homogeneity coefficient of the RQR series, measured using FAIC DIAG are within the acceptable limits specified by IEC 61267 and IAEA TRS-457. The RQR series (RQR 2 to RQR 8) are established as per IAEA TRS-457 specifications. Conclusions: The performance of the indegeneously developed FAICs was found satisfactory as per design and expectation and they qualify as absolute dosimeters for measurement of air-kerma of low energy regions and for direct calibration of field dosimeters and quality assurance test tools used for performance evaluation of diagnostic X-ray machines in the radiology department.

Authors

Institutions

Publication Details

Journal
Journal of Cancer Research and Therapeutics
Published
2024-10-01
DOI
https://doi.org/10.4103/jcrt.jcrt_2296_23
Citations
1
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.61
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of absolute dosimeter for standardization of low-energy synchrotron radiation and establishing of diagnostic X-ray beam qualities

Balvinder Kaur Sapra, Rahul Chaudhary, Sudhir Kumar, Sunil Dutt Sharma
1 citations
Journal of Cancer Research and Therapeutics
Advanced Radiotherapy Techniques
0.61
article

Development of absolute dosimeter for standardization of low-energy synchrotron radiation and establishing of diagnostic X-ray beam qualities

Balvinder Kaur Sapra, Rahul Chaudhary, Sudhir Kumar, Sunil Dutt Sharma
article en
1 citations

Abstract

ABSTRACT Purpose: Low energy X-rays from synchrotron radiation (SR) and X-ray generators are used for diagnostic imaging procedures. The dosimetry of low energy X-rays is challenging. For these types of radiations, the free-air ionization chamber (FAIC) is considered as the primary standard for absolute determination of air-kerma, which is an important quantity for the patient dose during diagnostic radiology investigation. Materials and Methods: To fulfil above requirement, a miniature parallel-plate FAIC (FAIC SR ) and portable parallel-plate FAIC (FAIC DIAG ) were designed and fabricated as the absolute dosimeter for standardization of SR beams and for establishing diagnostic X-rays beam qualities, RQR series respectively. Experiments were performed in SR beams (8-25 keV) as well as in diagnostic X-ray beams of 40, 50, 60, 70, 81, 90, 100 KVp for evaluating the basic characteristics of these chambers. Results: The basic characteristics of these chambers were in conformity with International Atomic Energy Agency (IAEA) TRS-374 and International Electrotechnical Commission (IEC) 61674, 60731. Dosimetric beam quality parameters such as the first half-value layer and homogeneity coefficient of the RQR series, measured using FAIC DIAG are within the acceptable limits specified by IEC 61267 and IAEA TRS-457. The RQR series (RQR 2 to RQR 8) are established as per IAEA TRS-457 specifications. Conclusions: The performance of the indegeneously developed FAICs was found satisfactory as per design and expectation and they qualify as absolute dosimeters for measurement of air-kerma of low energy regions and for direct calibration of field dosimeters and quality assurance test tools used for performance evaluation of diagnostic X-ray machines in the radiology department.

Journal of Cancer Research and TherapeuticsVol. 20(6)
Bhabha Atomic Research Centre (IN), Homi Bhabha National Institute (IN)
Openalex Percentile: Top 27%
Advanced Radiotherapy Techniques
0.61
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.