Improving Electromyography Electrode Placement Accuracy in Transtibial Amputees: A Comparative Study of Ultrasound and Palpation Methods

In the past decade, significant focus has been on electromyography (EMG) control of prostheses in transtibial amputees (TTAs). Reliable signal acquisition requires accurate EMG electrode placement. Conventional electrode placement methods are challenging due to altered post-surgical anatomy. This study investigated the application of ultrasound imaging for placement of EMG electrodes in TTAs. Four residual limb muscles, Tibialis Anterior (TA), Peroneus Longus (PL), Gastrocnemius Medial (GM), and Gastrocnemius Lateral (GL), were examined in 9 unilateral TTAs. Ultrasound was used to identify each muscle belly's thickest part and fiber orientation. A Certified Prosthetist Orthotist (CPO) then performed palpation to identify muscle bellies, blinded to ultrasound findings. Distances between ultrasound- and palpation-identified spots were measured. EMG data were contrasted between methods in terms of root mean square (RMS) amplitude and signal-to-noise ratio (SNR). The results indicated that Ultrasound-guided placement produced slightly higher, though non-significant, signal amplitudes (p = 0.06) and significantly higher SNR (p = 0.04). Moreover, palpation misidentified muscles in four cases. In 72.2% of cases, the distance between ultrasound- and palpation-identified spots was more than 10 mm. The mean distance was the greatest for PL and GL. Relying on palpation to identify PL and TA in TTAs may provide irrelevant EMG due to erroneous placement. Using ultrasound imaging can avoid this and, in addition to accurate muscle identification, may improve signal amplitude and SNR. In conclusion, ultrasound imaging is a valuable tool for enhancing the accuracy of EMG electrode placement in TTAs, which may lead to better prosthetic control outcomes.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Neural Systems and Rehabilitation Engineering
Published
2024-12-20
DOI
https://doi.org/10.1109/tnsre.2024.3520720
Citations
3
Primary Topic
Muscle activation and electromyography studies
Type
article
Field-Weighted Citation Impact
0.55
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Improving Electromyography Electrode Placement Accuracy in Transtibial Amputees: A Comparative Study of Ultrasound and Palpation Methods

Atli Örn Sverrisson, Kristín Briem, Sigurður Brynjólfsson, Faranak Rostamjoud et al.
3 citations
IEEE Transactions on Neural Systems and Rehabilitation Engineering
Muscle activation and electromyography studies
0.55
article

Improving Electromyography Electrode Placement Accuracy in Transtibial Amputees: A Comparative Study of Ultrasound and Palpation Methods

Atli Örn Sverrisson, Kristín Briem, Sigurður Brynjólfsson, Faranak Rostamjoud, Friðrika Björk Þorkelsdóttir
article en
3 citations

Abstract

In the past decade, significant focus has been on electromyography (EMG) control of prostheses in transtibial amputees (TTAs). Reliable signal acquisition requires accurate EMG electrode placement. Conventional electrode placement methods are challenging due to altered post-surgical anatomy. This study investigated the application of ultrasound imaging for placement of EMG electrodes in TTAs. Four residual limb muscles, Tibialis Anterior (TA), Peroneus Longus (PL), Gastrocnemius Medial (GM), and Gastrocnemius Lateral (GL), were examined in 9 unilateral TTAs. Ultrasound was used to identify each muscle belly's thickest part and fiber orientation. A Certified Prosthetist Orthotist (CPO) then performed palpation to identify muscle bellies, blinded to ultrasound findings. Distances between ultrasound- and palpation-identified spots were measured. EMG data were contrasted between methods in terms of root mean square (RMS) amplitude and signal-to-noise ratio (SNR). The results indicated that Ultrasound-guided placement produced slightly higher, though non-significant, signal amplitudes (p = 0.06) and significantly higher SNR (p = 0.04). Moreover, palpation misidentified muscles in four cases. In 72.2% of cases, the distance between ultrasound- and palpation-identified spots was more than 10 mm. The mean distance was the greatest for PL and GL. Relying on palpation to identify PL and TA in TTAs may provide irrelevant EMG due to erroneous placement. Using ultrasound imaging can avoid this and, in addition to accurate muscle identification, may improve signal amplitude and SNR. In conclusion, ultrasound imaging is a valuable tool for enhancing the accuracy of EMG electrode placement in TTAs, which may lead to better prosthetic control outcomes.

IEEE Transactions on Neural Systems and Rehabilitation EngineeringVol. 33
University of Iceland (IS)
Openalex Percentile: Top 36%
Muscle activation and electromyography studies
0.55
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.