Enhanced Stability and Sensitivity for CA-125 Detection Under Microfluidic Shear Flow Using Polyethylene Glycol-Coated Biosensor

The microfluidic-based point-of-care (POC) diagnostic tool has garnered significant interest in recent years, offering rapid and cost-effective disease detection. There is a growing trend toward integrating microfluidic platforms with biosensors, aligning lab-on-a-chip technologies with POC diagnostic devices. Despite numerous efforts to incorporate biosensors into microfluidic systems, researchers have performed very limited investigations on the stability of biomarker detection when biosensors operate under microfluidic shear flow conditions. Gold nanoparticles (AuNPs) are a widely employed material in capacitive biosensors for antibody immobilization and sensitivity enhancement. However, AuNPs have limitations in providing stable detection of biomarkers within microfluidic shear flow due to their agglomeration nature. This study addresses these limitations by employing 2 kDa polyethylene glycol (PEG) as an intermediate biofunctional layer to immobilize CA-125 antibodies on gold-interdigitated electrodes for the stable and accurate detection of CA-125 antigens. The stabilities and sensitivities of AuNPs and PEG-coated biosensors are evaluated under both static drop and microfluidic shear flow conditions for CA-125 antigen detection. The experimental results demonstrate a capacitive signal response (5660 pF at 10 kHz) 2.2 times higher using the PEG-coated biosensor than the signal (2551 pF at 10 kHz) measured by the AuNP-coated biosensor in the detection of CA-125 antigen-antibody conjugation under static drop conditions, indicating the higher sensitivity of the PEG-coated biosensor. Additionally, the PEG-coated biosensor exhibits better consistency for the CA-125 antigen detection between static drop and microfluidic shear flow conditions (Cp decrease in percentage (ΔCp%↓) = 2.9% at 10 kHz) compared to the electrical signals measured using the AuNP-coated biosensor (ΔCp%↓ = 32.4% at 10 kHz), which suggests that the PEG-coated biosensor demonstrates higher stability for CA-125 antigen detection under microfluidic shear flow conditions. With these significant improvements brought by the PEG-coated biosensor, especially under microfluidic conditions, a substantial hurdle in developing electrical biosensors for POC diagnostic applications has been overcome, expediting further advancements in the field.

Authors

Institutions

Publication Details

Journal
ACS Omega
Published
2024-12-20
DOI
https://doi.org/10.1021/acsomega.4c07596
Citations
7
Primary Topic
Microfluidic and Bio-sensing Technologies
Type
article
Field-Weighted Citation Impact
1.29

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhanced Stability and Sensitivity for CA-125 Detection Under Microfluidic Shear Flow Using Polyethylene Glycol-Coated Biosensor

Niladri Talukder, Yudong Wang, Ming Chang Lu, Xiao Tong et al.
7 citations
ACS Omega
Microfluidic and Bio-sensing Technologies
1.29
article

Enhanced Stability and Sensitivity for CA-125 Detection Under Microfluidic Shear Flow Using Polyethylene Glycol-Coated Biosensor

Niladri Talukder, Yudong Wang, Ming Chang Lu, Xiao Tong, Eon Soo Lee, Bharath Babu Nunna
article en
7 citations

Abstract

The microfluidic-based point-of-care (POC) diagnostic tool has garnered significant interest in recent years, offering rapid and cost-effective disease detection. There is a growing trend toward integrating microfluidic platforms with biosensors, aligning lab-on-a-chip technologies with POC diagnostic devices. Despite numerous efforts to incorporate biosensors into microfluidic systems, researchers have performed very limited investigations on the stability of biomarker detection when biosensors operate under microfluidic shear flow conditions. Gold nanoparticles (AuNPs) are a widely employed material in capacitive biosensors for antibody immobilization and sensitivity enhancement. However, AuNPs have limitations in providing stable detection of biomarkers within microfluidic shear flow due to their agglomeration nature. This study addresses these limitations by employing 2 kDa polyethylene glycol (PEG) as an intermediate biofunctional layer to immobilize CA-125 antibodies on gold-interdigitated electrodes for the stable and accurate detection of CA-125 antigens. The stabilities and sensitivities of AuNPs and PEG-coated biosensors are evaluated under both static drop and microfluidic shear flow conditions for CA-125 antigen detection. The experimental results demonstrate a capacitive signal response (5660 pF at 10 kHz) 2.2 times higher using the PEG-coated biosensor than the signal (2551 pF at 10 kHz) measured by the AuNP-coated biosensor in the detection of CA-125 antigen-antibody conjugation under static drop conditions, indicating the higher sensitivity of the PEG-coated biosensor. Additionally, the PEG-coated biosensor exhibits better consistency for the CA-125 antigen detection between static drop and microfluidic shear flow conditions (Cp decrease in percentage (ΔCp%↓) = 2.9% at 10 kHz) compared to the electrical signals measured using the AuNP-coated biosensor (ΔCp%↓ = 32.4% at 10 kHz), which suggests that the PEG-coated biosensor demonstrates higher stability for CA-125 antigen detection under microfluidic shear flow conditions. With these significant improvements brought by the PEG-coated biosensor, especially under microfluidic conditions, a substantial hurdle in developing electrical biosensors for POC diagnostic applications has been overcome, expediting further advancements in the field.

ACS OmegaVol. 10(1)
New Jersey Institute of Technology (US), Weber State University (US), Brigham and Women's Hospital (US), Harvard University (US), Brookhaven National Laboratory (US), Advanced Energy Systems (United States) (US)
National Science Foundation, U.S. Department of Energy, Nanjing Institute of Technology, Office of Science, Translational Impacts, Brookhaven National Laboratory
Affordable and clean energy
Openalex Percentile: Top 21%
Microfluidic and Bio-sensing Technologies
1.29
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.