Accelerated Aging Effect on the Stability of the 3D-Printed Biodegradable Implant for Bone Defect Repairs

This article presents an evaluation of the accelerated aging impact on the structural properties of biodegradable PLA/HAp implants produced using 3D printing technology for use in traumatic bone defect repairs in individual patients. The designed biodegradable implants were sterilized with a radiation dose of 25 ± 0.99% kGy, then exposed to an accelerated aging process. Selected physicomechanical and chemical properties of biodegradable implants were evaluated with FT-IR spectra analyses and DSC. The accelerated aging process, carried out according to the ASTM F 1980:2002 "Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices", simulates three years of implant usage. It confirmed the stability of structural, physical and mechanical properties and proved the effectiveness and safety of the implants' application. The present study was conducted to determine the shelf-life of newly developed biodegradable implants proposed for the treatment of children and adolescents where bone growth still occurs by using accelerated aging methodologies, allowing the assessment of changes in performance that do not result in a negative impact on the safety of the medical device.

Authors

Institutions

Publication Details

Journal
Materials
Published
2024-12-19
DOI
https://doi.org/10.3390/ma17246218
Citations
1
Primary Topic
Bone Tissue Engineering Materials
Type
article
Field-Weighted Citation Impact
0.18

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accelerated Aging Effect on the Stability of the 3D-Printed Biodegradable Implant for Bone Defect Repairs

Agnieszka Kucharska-Jastrząbek, Edyta Chmal-Fudali, Agnieszka Gutowska, Bożena Wilbik-Hałgas et al.
1 citations
Materials
Bone Tissue Engineering Materials
0.18
article

Accelerated Aging Effect on the Stability of the 3D-Printed Biodegradable Implant for Bone Defect Repairs

Agnieszka Kucharska-Jastrząbek, Edyta Chmal-Fudali, Agnieszka Gutowska, Bożena Wilbik-Hałgas, Marcin H. Struszczyk, Paweł Kubiak, Katarzyna Kośla
article en
1 citations

Abstract

This article presents an evaluation of the accelerated aging impact on the structural properties of biodegradable PLA/HAp implants produced using 3D printing technology for use in traumatic bone defect repairs in individual patients. The designed biodegradable implants were sterilized with a radiation dose of 25 ± 0.99% kGy, then exposed to an accelerated aging process. Selected physicomechanical and chemical properties of biodegradable implants were evaluated with FT-IR spectra analyses and DSC. The accelerated aging process, carried out according to the ASTM F 1980:2002 "Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices", simulates three years of implant usage. It confirmed the stability of structural, physical and mechanical properties and proved the effectiveness and safety of the implants' application. The present study was conducted to determine the shelf-life of newly developed biodegradable implants proposed for the treatment of children and adolescents where bone growth still occurs by using accelerated aging methodologies, allowing the assessment of changes in performance that do not result in a negative impact on the safety of the medical device.

MaterialsVol. 17(24)
Institute of Security Technologies MORATEX (PL)
Narodowe Centrum Badań i Rozwoju
Openalex Percentile: Top 49%
Bone Tissue Engineering Materials
0.18
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.