Comparative Study of Fluence Distribution and Point Dose Using Arc-check and Delta4 Phantoms

The study aims to assess the fluence distribution and point dosage between two phantoms for patient-specific quality assurance on the Tomotherapy system. This was a retrospective study conducted on 15 patients who had radiation using the Helical Tomotherapy Machine (Radixact, Accuray Inc.). We used two phantoms to quantify the fluence produced by the treatment planning system (TPS) and recorded from the machine. The ArcCHECK (Sun-Nuclear) has 1386 diodes placed in a cylindrical configuration. The minimal resolution for this was 7 mm. The second was Delta<sup>4</sup>, supplied by ScandiDos. It has 1069 diode detectors arrayed in a crossed orthogonal configuration with a minimum resolution of 5 mm. All patient plans were transferred to these phantoms to validate the accuracy of treatment plan delivery. We used SunCHECK and ScandiDos Delta<sup>4</sup> software to compare the fluence produced by the TPS with the fluence measured by the equipment. In ArcCHECK, we used an external ionization chamber, cc13 (IBA dosimetry), whereas in Delta<sup>4</sup>,we employed a central diode detector to quantify point dosage. The mean and standard deviation of the gamma pass percentage with ArcCHECK were 98.3 ± 0.8%, with an average point dose deviation of ± 0.94%. The mean and standard deviation of the gamma pass percentage using Delta<sup>4</sup> was 99.1 ± 1.6%, while the average point dose deviation was ± 0.60%, both of which were well within the 3% tolerance employing the two phantoms.

Authors

Institutions

Publication Details

Journal
Journal of Medical Physics
Published
2024-10-01
DOI
https://doi.org/10.4103/jmp.jmp_130_24
Citations
8
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
4.86
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Study of Fluence Distribution and Point Dose Using Arc-check and Delta4 Phantoms

Benoy Kumar Singh, Munish Gairola, Sandeep Kumar Singh, Abhay Kumar Singh et al.
8 citations
Journal of Medical Physics
Advanced Radiotherapy Techniques
4.86
article

Comparative Study of Fluence Distribution and Point Dose Using Arc-check and Delta4 Phantoms

Benoy Kumar Singh, Munish Gairola, Sandeep Kumar Singh, Abhay Kumar Singh, Manindra Bhushan, Anuj Kumar, Vikram Vikram, Dipesh
article en
8 citations

Abstract

The study aims to assess the fluence distribution and point dosage between two phantoms for patient-specific quality assurance on the Tomotherapy system. This was a retrospective study conducted on 15 patients who had radiation using the Helical Tomotherapy Machine (Radixact, Accuray Inc.). We used two phantoms to quantify the fluence produced by the treatment planning system (TPS) and recorded from the machine. The ArcCHECK (Sun-Nuclear) has 1386 diodes placed in a cylindrical configuration. The minimal resolution for this was 7 mm. The second was Delta<sup>4</sup>, supplied by ScandiDos. It has 1069 diode detectors arrayed in a crossed orthogonal configuration with a minimum resolution of 5 mm. All patient plans were transferred to these phantoms to validate the accuracy of treatment plan delivery. We used SunCHECK and ScandiDos Delta<sup>4</sup> software to compare the fluence produced by the TPS with the fluence measured by the equipment. In ArcCHECK, we used an external ionization chamber, cc13 (IBA dosimetry), whereas in Delta<sup>4</sup>,we employed a central diode detector to quantify point dosage. The mean and standard deviation of the gamma pass percentage with ArcCHECK were 98.3 ± 0.8%, with an average point dose deviation of ± 0.94%. The mean and standard deviation of the gamma pass percentage using Delta<sup>4</sup> was 99.1 ± 1.6%, while the average point dose deviation was ± 0.60%, both of which were well within the 3% tolerance employing the two phantoms.

Journal of Medical PhysicsVol. 49(4)
Lala Lajpat Rai Memorial Medical College (IN), Rajiv Gandhi Cancer Institute and Research Centre (IN), BLK Super Speciality Hospital (IN), GLA University (IN)
Openalex Percentile: Top 4%
Advanced Radiotherapy Techniques
4.86
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Comparative Study of Fluence Distribution and Point Dose Using Arc-check and Delta4 Phantoms — Benoy Kumar Singh, Munish Gairola, et al. · Journal of Medical Physics (2024) | TGRS Research Map | TGRS