Dosimetric comparison of M6 CyberKnife plans optimized with Precision and RayStation 12A treatment planning systems
RS is an adequate alternative for CK planning as it is possible to obtain plan quality comparable to Precision. However, the optimization time is longer compared to Precision and more attention must be paid to the choice of the initial conditions in terms of the number of beams and nodes.
Authors
- Raphaël Moeckli (ORCID: https://orcid.org/0000-0002-5885-934X)
- Véronique Vallet (ORCID: https://orcid.org/0000-0002-1876-4805)
- François Bochud (ORCID: https://orcid.org/0000-0003-2076-0296)
- Jean Bourhis (ORCID: https://orcid.org/0000-0001-5162-1171)
- Maude Gondré (ORCID: https://orcid.org/0009-0000-4736-7799)
Institutions
- Institute of Radiation Physics (CH)
- University of Lausanne (CH)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2024-12-19
- DOI
- https://doi.org/10.1002/acm2.14585
- Citations
- 1
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.61