Operational Verification of a Parallel Open/Closeable Forceps Tip Mechanism for Forceps-Type Mini-PET

Forceps-type mini-positron emission tomography (mini-PET) has been proposed as an intraoperative device for examining metastatic lymph nodes in the treatment of esophageal cancer. Although this forceps-type mini-PET detects cancer by radiation measurement, the scissor-like tip of the device affects the measurement accuracy. Therefore, to improve the detection sensitivity of the forceps-type mini-PET without depending on the operator, we fabricated a forceps tip mechanism using a parallel-link for parallel opening and closing motions, and verified the parallel motion and the force required for opening and closing. Consequently, we confirmed parallel motion from the opening and closing widths of the tip detectors. In addition, we confirmed that the forces required for opening and closing were significantly smaller than those required for the conventional mechanism.

Authors

Institutions

Publication Details

Journal
Journal of Robotics and Mechatronics
Published
2024-12-19
DOI
https://doi.org/10.20965/jrm.2024.p1550
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Operational Verification of a Parallel Open/Closeable Forceps Tip Mechanism for Forceps-Type Mini-PET

Kazuya Kawamura, Taiga Yamaya, M. Takahashi, Shigeki Ito et al.
Journal of Robotics and Mechatronics
Medical Imaging Techniques and Applications
article

Operational Verification of a Parallel Open/Closeable Forceps Tip Mechanism for Forceps-Type Mini-PET

Kazuya Kawamura, Taiga Yamaya, M. Takahashi, Shigeki Ito, Hiroto Hayashi
article en

Abstract

Forceps-type mini-positron emission tomography (mini-PET) has been proposed as an intraoperative device for examining metastatic lymph nodes in the treatment of esophageal cancer. Although this forceps-type mini-PET detects cancer by radiation measurement, the scissor-like tip of the device affects the measurement accuracy. Therefore, to improve the detection sensitivity of the forceps-type mini-PET without depending on the operator, we fabricated a forceps tip mechanism using a parallel-link for parallel opening and closing motions, and verified the parallel motion and the force required for opening and closing. Consequently, we confirmed parallel motion from the opening and closing widths of the tip detectors. In addition, we confirmed that the forces required for opening and closing were significantly smaller than those required for the conventional mechanism.

Journal of Robotics and MechatronicsVol. 36(6)
Chiba University (JP), National Institutes for Quantum Science and Technology (JP), Mirai Hospital (JP)
Japan Agency for Medical Research and Development, Nakatani Foundation for Advancement of Measuring Technologies in Biomedical Engineering
Openalex Percentile: Top 60%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.