Sample Thickness Effect on Quantitative Analysis by Hand‐Held X‐Ray Fluorescence Spectrometers

ABSTRACT Commercially available hand‐held X‐ray fluorescence (HH‐XRF) spectrometers have their own quantitative analysis software. When the elements Se and Cd in soil samples are measured and quantified, the calculated concentrations sometimes fail by a factor of 1/2 for Cd, although the values obtained for Se are accurate. This is because of the escape depth length of fluorescent X‐rays. Easy ways to overcome this problem are presented; two point calibration, or at least one point and one blank is enough. Problems associated with mercury evaporation from the sample is also discussed. Filter paper analysis is used as an example for the present work, and a five‐filter pile of ash‐less paper is concluded to have better sensitivity than a single paper filter analysis.

Authors

Institutions

Publication Details

Journal
X-Ray Spectrometry
Published
2024-12-17
DOI
https://doi.org/10.1002/xrs.3471
Citations
1
Primary Topic
X-ray Spectroscopy and Fluorescence Analysis
Type
article
Field-Weighted Citation Impact
0.61

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Sample Thickness Effect on Quantitative Analysis by Hand‐Held X‐Ray Fluorescence Spectrometers

Damdinsuren Bolortuya, Shinichi Ishizuka, Hussain Alawadhi, Jun Kawai et al.
1 citations
X-Ray Spectrometry
X-ray Spectroscopy and Fluorescence Analysis
0.61
article

Sample Thickness Effect on Quantitative Analysis by Hand‐Held X‐Ray Fluorescence Spectrometers

Damdinsuren Bolortuya, Shinichi Ishizuka, Hussain Alawadhi, Jun Kawai, M. Nishimoto, Mohamed Shameer
article en
1 citations

Abstract

ABSTRACT Commercially available hand‐held X‐ray fluorescence (HH‐XRF) spectrometers have their own quantitative analysis software. When the elements Se and Cd in soil samples are measured and quantified, the calculated concentrations sometimes fail by a factor of 1/2 for Cd, although the values obtained for Se are accurate. This is because of the escape depth length of fluorescent X‐rays. Easy ways to overcome this problem are presented; two point calibration, or at least one point and one blank is enough. Problems associated with mercury evaporation from the sample is also discussed. Filter paper analysis is used as an example for the present work, and a five‐filter pile of ash‐less paper is concluded to have better sensitivity than a single paper filter analysis.

X-Ray SpectrometryVol. 54(4)
Ryukoku University (JP), Kyoto University (JP), University of Sharjah (AE), National University of Mongolia (MN)
Ministry of Education, Culture, Sports, Science and Technology, Japan International Cooperation Agency, Ryukoku University
Life in Land
Openalex Percentile: Top 28%
X-ray Spectroscopy and Fluorescence Analysis
0.61
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Sample Thickness Effect on Quantitative Analysis by Hand‐Held X‐Ray Fluorescence Spectrometers — Damdinsuren Bolortuya, Shinichi Ishizuka, et al. · X-Ray Spectrometry (2024) | TGRS Research Map | TGRS