Design and Characterisation of an EIT Voltage Conditioning Module

Abstract Electrical Impedance Tomography is used to image the cross-sectional conductivity of an object. It is clinically used for high-frame-rate imaging of lung ventilation. Most current systems use very limited current injection waveforms and patterns. We have developed a flexible system for researching alternatives. This paper covers our voltage conditioning module’s design and characterisation. The device was shown to work up to the desired 10 MHz. The responses were uniform with minor gain mismatch between all channels after manufacturing variations. The low-frequency (<1 MHz) crosstalk is on the order -55 dB, and the worst cases around 8 MHz are around -35 dB.

Authors

Institutions

Publication Details

Journal
Current Directions in Biomedical Engineering
Published
2024-12-01
DOI
https://doi.org/10.1515/cdbme-2024-2165
Primary Topic
Non-Destructive Testing Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and Characterisation of an EIT Voltage Conditioning Module

Jack Wilkie, Alberto Battistel, Rongqing Chen, Knut Möller
Current Directions in Biomedical Engineering
Non-Destructive Testing Techniques
article

Design and Characterisation of an EIT Voltage Conditioning Module

Jack Wilkie, Alberto Battistel, Rongqing Chen, Knut Möller
article en

Abstract

Abstract Electrical Impedance Tomography is used to image the cross-sectional conductivity of an object. It is clinically used for high-frame-rate imaging of lung ventilation. Most current systems use very limited current injection waveforms and patterns. We have developed a flexible system for researching alternatives. This paper covers our voltage conditioning module’s design and characterisation. The device was shown to work up to the desired 10 MHz. The responses were uniform with minor gain mismatch between all channels after manufacturing variations. The low-frequency (<1 MHz) crosstalk is on the order -55 dB, and the worst cases around 8 MHz are around -35 dB.

Current Directions in Biomedical EngineeringVol. 10(4)
Furtwangen University (DE)
Openalex Percentile: Top 76%
Non-Destructive Testing Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.