Design and Characterisation of an EIT Current Supply Module

Abstract Electrical Impedance Tomography is used to image the cross-sectional conductivity of an object. It is clinically used for high-frame-rate imaging of lung ventilation. Most current systems use very limited current injection wave-forms and patterns. We have developed a flexible system for researching alternatives. This paper covers our current supply module's design and characterisation. We tested the input filtering, Bandwidth, and parasitic capacitance. The imput fitering functioned similarly to the simulation of the design. However, the bandwich was limited to approximately 1 MHz instead of the desired 10 MHz due to excessive parasitic capacitance. We proposed points to redesign that should reduce the capacitance by a factor of 10 and correspondingly increase dandwidth tenfold to 10 MHz.

Authors

Institutions

Publication Details

Journal
Current Directions in Biomedical Engineering
Published
2024-12-01
DOI
https://doi.org/10.1515/cdbme-2024-2164
Primary Topic
Non-Destructive Testing Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and Characterisation of an EIT Current Supply Module

Jack Wilkie, Alberto Battistel, Chen Rong-qin, Knut Möller
2 links
Current Directions in Biomedical Engineering
Non-Destructive Testing Techniques
article

Design and Characterisation of an EIT Current Supply Module

Jack Wilkie, Alberto Battistel, Chen Rong-qin, Knut Möller
article en

Abstract

Abstract Electrical Impedance Tomography is used to image the cross-sectional conductivity of an object. It is clinically used for high-frame-rate imaging of lung ventilation. Most current systems use very limited current injection wave-forms and patterns. We have developed a flexible system for researching alternatives. This paper covers our current supply module's design and characterisation. We tested the input filtering, Bandwidth, and parasitic capacitance. The imput fitering functioned similarly to the simulation of the design. However, the bandwich was limited to approximately 1 MHz instead of the desired 10 MHz due to excessive parasitic capacitance. We proposed points to redesign that should reduce the capacitance by a factor of 10 and correspondingly increase dandwidth tenfold to 10 MHz.

Current Directions in Biomedical EngineeringVol. 10(4)
Furtwangen University (DE)
Openalex Percentile: Top 76%
Non-Destructive Testing Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.