Development of a model system for in vitro simulated application testing of neurovascular stent systems

Abstract In vitro test setups for simulated application testing are necessary for physician training as well as for development and benchmarking of various devices in interventional neuroradiology. Within the current study, we implemented a test setup for in vitro simulated application testing. Trackability in the distal catheter area, which was positioned inside a complex 3D vascular model was analyzed for two neurovascular stent systems (Neuroform EZ 4.0x20 mm and Enterprise VRD 4.5x14 mm) combined with different microcatheters (Excelsior XT-27 Flex, PX Slim Delivery Microcatheter and Prowler Select Plus Infusion Catheter). The described test setup allows for a diffentiation of neurovascular stent systems with regard to the trackability force as a major procedure related indicator for device handling. Therefore, the simulated application system presented within the current work represents an indispensable tool for safely and efficiently developing and evaluating new technologies for interventional neuroradiology prior to clinical implementation.

Authors

Institutions

Publication Details

Journal
Current Directions in Biomedical Engineering
Published
2024-12-01
DOI
https://doi.org/10.1515/cdbme-2024-2145
Primary Topic
Cerebrovascular and Carotid Artery Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a model system for in vitro simulated application testing of neurovascular stent systems

Christoph Brandt-Wunderlich, W. J. Schmidt, Marc‐André Weber, Felix Streckenbach et al.
Current Directions in Biomedical Engineering
Cerebrovascular and Carotid Artery Diseases
article

Development of a model system for in vitro simulated application testing of neurovascular stent systems

Christoph Brandt-Wunderlich, W. J. Schmidt, Marc‐André Weber, Felix Streckenbach, Sönke Langner, Stefan Siewert, Michael Stiehm, Daniel Cantré, Hagen Paetow, Klaus‐Peter Schmitz
article en

Abstract

Abstract In vitro test setups for simulated application testing are necessary for physician training as well as for development and benchmarking of various devices in interventional neuroradiology. Within the current study, we implemented a test setup for in vitro simulated application testing. Trackability in the distal catheter area, which was positioned inside a complex 3D vascular model was analyzed for two neurovascular stent systems (Neuroform EZ 4.0x20 mm and Enterprise VRD 4.5x14 mm) combined with different microcatheters (Excelsior XT-27 Flex, PX Slim Delivery Microcatheter and Prowler Select Plus Infusion Catheter). The described test setup allows for a diffentiation of neurovascular stent systems with regard to the trackability force as a major procedure related indicator for device handling. Therefore, the simulated application system presented within the current work represents an indispensable tool for safely and efficiently developing and evaluating new technologies for interventional neuroradiology prior to clinical implementation.

Current Directions in Biomedical EngineeringVol. 10(4)
Institute for ImplantTechnology and Biomaterials (DE), University of Rostock (DE)
Openalex Percentile: Top 57%
Cerebrovascular and Carotid Artery Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.