Reducing artifacts in electrically evoked auditory potentials (eAEP)

Abstract Introduction: In auditory evoked potentials (= AEP), auditory brainstem response (= ABR) is the gold standard for objective analysis of the auditory pathway using acoustical stimulators like headphones during hearing screening and patient assessment. Where no response can be recorded, electrical stimulation can be used in pre-, intra-, or postoperative scenarios to confirm the excitability of the auditory nerve (= eAEP). In eAEP, the recording of brainstem response (= eABR) is mostly the method of choice. Sometimes in eABR, the stimulation artifact can not only harden but even prevent the detection of the response waveforms. Methods: Many factors can influence the stimulation artifact when recording an eAEP. Neurological recording systems are recommended as well as some approved AEP systems traditionally used in ENT for amplification to avoid large stimulation artifacts in the response waveforms, but able to recover in <1 ms from saturation. High stimulation rates (~90 Hz) can lead to overlapping of the response waveforms with the following stimulation artifact. In preoperative eAEP, stimulation artifacts are often high due to extra-cochlear stimulation. In intra- and postoperative eAEP, the artifact is lower and mostly similar across patients. Results: The recommended recording systems show low stimulation artifacts in almost any case for intra- and postoperative eABR. In contrast, later responses, like electrically evoked auditory middle-latency response (= eAMLR) and electrically evoked auditory late response (= eALR), are less affected by artifacts. Conclusion: The stimulation artifact can be a problematic factor. Depending on the EP system, intra- and postoperative eABR can be challenging when trying to record the response waves. Where possible, clinicians should think about eAMLR or eALR for results with less stimulation artifact. Finally, the most important factor remains using a recording system that can handle the stimulation artifact. We do not recommend recording devices with high artifacts for preoperative eABR. For intra- and postoperative eABR, the recommended systems can prevent challenging recording.

Authors

Institutions

Publication Details

Journal
Current Directions in Biomedical Engineering
Published
2024-12-01
DOI
https://doi.org/10.1515/cdbme-2024-2124
Primary Topic
Neural Networks and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reducing artifacts in electrically evoked auditory potentials (eAEP)

Daniel Polterauer, Giacomo Mandruzzato, Marek Polák, Maike Neuling et al.
Current Directions in Biomedical Engineering
Neural Networks and Applications
article

Reducing artifacts in electrically evoked auditory potentials (eAEP)

Daniel Polterauer, Giacomo Mandruzzato, Marek Polák, Maike Neuling, John‐Martin Hempel, Joachim Müller
article en

Abstract

Abstract Introduction: In auditory evoked potentials (= AEP), auditory brainstem response (= ABR) is the gold standard for objective analysis of the auditory pathway using acoustical stimulators like headphones during hearing screening and patient assessment. Where no response can be recorded, electrical stimulation can be used in pre-, intra-, or postoperative scenarios to confirm the excitability of the auditory nerve (= eAEP). In eAEP, the recording of brainstem response (= eABR) is mostly the method of choice. Sometimes in eABR, the stimulation artifact can not only harden but even prevent the detection of the response waveforms. Methods: Many factors can influence the stimulation artifact when recording an eAEP. Neurological recording systems are recommended as well as some approved AEP systems traditionally used in ENT for amplification to avoid large stimulation artifacts in the response waveforms, but able to recover in <1 ms from saturation. High stimulation rates (~90 Hz) can lead to overlapping of the response waveforms with the following stimulation artifact. In preoperative eAEP, stimulation artifacts are often high due to extra-cochlear stimulation. In intra- and postoperative eAEP, the artifact is lower and mostly similar across patients. Results: The recommended recording systems show low stimulation artifacts in almost any case for intra- and postoperative eABR. In contrast, later responses, like electrically evoked auditory middle-latency response (= eAMLR) and electrically evoked auditory late response (= eALR), are less affected by artifacts. Conclusion: The stimulation artifact can be a problematic factor. Depending on the EP system, intra- and postoperative eABR can be challenging when trying to record the response waves. Where possible, clinicians should think about eAMLR or eALR for results with less stimulation artifact. Finally, the most important factor remains using a recording system that can handle the stimulation artifact. We do not recommend recording devices with high artifacts for preoperative eABR. For intra- and postoperative eABR, the recommended systems can prevent challenging recording.

Current Directions in Biomedical EngineeringVol. 10(4)
LMU Klinikum (DE), MED-EL (Austria) (AT), Ludwig-Maximilians-Universität München (DE)
Openalex Percentile: Top 79%
Neural Networks and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.