Force signals time alignment using cross-correlation

Calibration of force in measuring systems where the force is time dependent i.e. dynamically changing, introduces additional complexities versus quasi-static calibration conditions. The basic need in the case of time variable force signals is to synchronise the measurement series from unit under test and the measurement series from the reference standard. As they are typically acquired by different measuring systems, post-processing is generally necessary to align the measured samples to try to eliminate the influence of the sample misalignment from the resulting testing machine indication error. The paper focuses on the time alignment of measurement series at testing machine calibration with cyclic force excitation using cross-correlation to determine the necessary time correction. The standard deviation of the error distribution for the two measured signals after time correction was below 0.2 % for 5 kN ± 1 kN cycling force for 1 Hz, 5 Hz, and 10 Hz cycling frequency.

Authors

Institutions

Publication Details

Journal
Measurement Sensors
Published
2024-12-18
DOI
https://doi.org/10.1016/j.measen.2024.101375
Citations
2
Primary Topic
Speech and Audio Processing
Type
article
Field-Weighted Citation Impact
0.70
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Force signals time alignment using cross-correlation

Miha Hiti
2 citations
Measurement Sensors
Speech and Audio Processing
0.70
article

Force signals time alignment using cross-correlation

Miha Hiti
article en
2 citations

Abstract

Calibration of force in measuring systems where the force is time dependent i.e. dynamically changing, introduces additional complexities versus quasi-static calibration conditions. The basic need in the case of time variable force signals is to synchronise the measurement series from unit under test and the measurement series from the reference standard. As they are typically acquired by different measuring systems, post-processing is generally necessary to align the measured samples to try to eliminate the influence of the sample misalignment from the resulting testing machine indication error. The paper focuses on the time alignment of measurement series at testing machine calibration with cyclic force excitation using cross-correlation to determine the necessary time correction. The standard deviation of the error distribution for the two measured signals after time correction was below 0.2 % for 5 kN ± 1 kN cycling force for 1 Hz, 5 Hz, and 10 Hz cycling frequency.

Measurement SensorsVol. 38
Slovenian National Building and Civil Engineering Institute (SI)
Openalex Percentile: Top 29%
Speech and Audio Processing
0.70
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.