Study on the 30 MN force calibration system in tension with failure protection

The maximum risk in the calibration of tension force in the meganewton range comes from failure of the transducer. The impact caused either by the yield fracture or by the unforeseen brittle fracture can lead personal injury or equipment failure. In this work the deformation and energy effect of the 30 MN transducer upon fracture was evaluated. A new tension force calibration system with a failure protection function witch is consisted of micro-displacement monitoring devices and thin shell structure was built. The thin shell structure is made of stainless steel with multiple arrays honeycomb bulges that will produce large deformation to absorb energy upon impact. The dual ball nuts structure conforming to the Annex A.4 of ISO 376 requirements is applicable for this system, and the coaxial can be guaranteed in calibration. Measurements were performed at the 60 MN build-up machine (BM) in FJIM. It is shown that the system has sufficient stiffness with a 30 MN force, and the participation of the tension system did not influence the measurements in BM. Displacement at 0.02 mm level can be monitored in real time and deformation curve to warn yield failure proactively, and the passive protection structure absorbed energy in an unexpected brittle fracture at nearly 18 MN and protected the BM and personnel safety.

Authors

Institutions

Publication Details

Journal
Measurement Sensors
Published
2024-12-18
DOI
https://doi.org/10.1016/j.measen.2024.101369
Primary Topic
Mechanical stress and fatigue analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Study on the 30 MN force calibration system in tension with failure protection

Xie Jie, Yong Zhang, Shuo Lin, Chi Hui et al.
3 links
Measurement Sensors
Mechanical stress and fatigue analysis
article

Study on the 30 MN force calibration system in tension with failure protection

Xie Jie, Yong Zhang, Shuo Lin, Chi Hui, Que Pengfeng
article en

Abstract

The maximum risk in the calibration of tension force in the meganewton range comes from failure of the transducer. The impact caused either by the yield fracture or by the unforeseen brittle fracture can lead personal injury or equipment failure. In this work the deformation and energy effect of the 30 MN transducer upon fracture was evaluated. A new tension force calibration system with a failure protection function witch is consisted of micro-displacement monitoring devices and thin shell structure was built. The thin shell structure is made of stainless steel with multiple arrays honeycomb bulges that will produce large deformation to absorb energy upon impact. The dual ball nuts structure conforming to the Annex A.4 of ISO 376 requirements is applicable for this system, and the coaxial can be guaranteed in calibration. Measurements were performed at the 60 MN build-up machine (BM) in FJIM. It is shown that the system has sufficient stiffness with a 30 MN force, and the participation of the tension system did not influence the measurements in BM. Displacement at 0.02 mm level can be monitored in real time and deformation curve to warn yield failure proactively, and the passive protection structure absorbed energy in an unexpected brittle fracture at nearly 18 MN and protected the BM and personnel safety.

Measurement SensorsVol. 38
Fujian Metrology Institute (CN)
Openalex Percentile: Top 74%
Mechanical stress and fatigue analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Study on the 30 MN force calibration system in tension with failure protection — Xie Jie, Yong Zhang, et al. · Measurement Sensors (2024) | TGRS Research Map | TGRS