An Improved Setup for the Realisation of UTC(SCL) in the Standards and Calibration Laboratory
This paper describes an improved time scale system setup at the Standards and Calibration Laboratory (SCL) for the local realization of UTC, UTC(SCL). The original setup was based on the caesium frequency standards and a microphase stepper. An active hydrogen maser has been recently integrated into the system to provide an additional short-term stable frequency reference, and a frequency steering algorithm has been developed to calculate the microphase stepper settings to improve the accuracy and stability for the UTC(SCL).
Authors
- C. F. Au Yeung
- S. L. Yang
Institutions
- Food and Health Bureau (CN)
- Standards and Calibration Laboratory
Publication Details
- Published
- 2024-09-22
- DOI
- https://doi.org/10.1109/uffc-js60046.2024.10793811
- Primary Topic
- Advanced Frequency and Time Standards
- Type
- article
- Field-Weighted Citation Impact
- 0.00