Optimization of Beamforming Strategy Applied to a Dual Core Probe for US-Guided Sonoporation
In this work, a new ultrasound probe with dual-mode capabilities is proposed for ultrasound-guided sonoporation. This probe combines a high-frequency linear array (20 MHz) for the imaging function and a low-frequency matrix array (1MHz with 256 elements) for the therapy function. This work focuses on the low-frequency part of the probe. Different beamforming are studied to meet the sonoporation requirements. In-vitro tests are undertaken to validate one of the chosen beamforming and to validate the probe properties for sonoporation applications.
Authors
- Franck Levassort (ORCID: https://orcid.org/0000-0002-2019-9390)
- Rémi Rouffaud (ORCID: https://orcid.org/0000-0003-2752-5459)
- Jean‐Michel Escoffre (ORCID: https://orcid.org/0000-0003-1041-6950)
- Mathieu Legros (ORCID: https://orcid.org/0000-0003-3807-8594)
- Marie M. Le Roy (ORCID: https://orcid.org/0000-0003-1063-4050)
- Dominique Certon (ORCID: https://orcid.org/0000-0001-5963-7119)
- Damien Fouan (ORCID: https://orcid.org/0000-0002-7120-7223)
- Juline Cloet
Institutions
- Inserm (FR)
- Vermon (France) (FR)
Publication Details
- Published
- 2024-09-22
- DOI
- https://doi.org/10.1109/uffc-js60046.2024.10793918
- Primary Topic
- Ultrasound Imaging and Elastography
- Type
- article
- Field-Weighted Citation Impact
- 0.00