Microwave Glucose Sensing Using Double Circular Split Ring Resonators for Improved Sensitivity: The Role of Artificial Blood Plasma and Deionized Water

In 2021, approximately 537 million people were diagnosed with diabetes mellitus. With rates expected to rise, health expenditures are projected to reach one trillion USD by 2030. Thus, measuring glucose levels is essential for rationalizing the costs of public health systems. In this context, this article presents two major contributions. First, it demonstrates that using deionized water (DI-water) as a reference for glucose sensing is not a reliable approach for representing human blood plasma (BP), as it lacks ions and suppresses essential effects such as losses. As an alternative, we investigate the use of an artificial BP solution (ABPS) that closely resembles real human BP. Characterized over a range from 500 MHz to 10 GHz, ABPS shows marginal differences in real permittivity but significant differences in imaginary permittivity compared to DI-water. The second contribution is the design of a highly sensitive microwave (MW) sensor based on double concentric circular split ring resonator (DCCSRR) on a <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$49\times 45$ </tex-math></inline-formula> mm ROGERS 5880 TM substrate. This sensor can differentiate glucose concentrations from 0 to 400 mg/dL, exceeding the relevant range for diabetic individuals (50–300 mg/dL). The DCCSRR operates at 2.48 GHz and can detect minimal concentration variations of 25 mg/dL in low concentrations, representing a significant advancement in the field. Different from most sensitive approaches available to date, this structure operates in a nonlicensed band and a fully passive form, offering flexibility for implementation and low cost. These characteristics position it as a state-of-the-art solution in MW glucose sensors.

Authors

Institutions

Publication Details

Journal
IEEE Sensors Journal
Published
2024-12-18
DOI
https://doi.org/10.1109/jsen.2024.3516571
Citations
7
Primary Topic
Wireless Body Area Networks
Type
article
Field-Weighted Citation Impact
1.29

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Microwave Glucose Sensing Using Double Circular Split Ring Resonators for Improved Sensitivity: The Role of Artificial Blood Plasma and Deionized Water

Mateus I. O. Souza, Vinícius M. Pepino, J. Alarcon, Natália M. Santos et al.
7 citations
IEEE Sensors Journal
Wireless Body Area Networks
1.29
article

Microwave Glucose Sensing Using Double Circular Split Ring Resonators for Improved Sensitivity: The Role of Artificial Blood Plasma and Deionized Water

Mateus I. O. Souza, Vinícius M. Pepino, J. Alarcon, Natália M. Santos, Ben‐Hur V. Borges, Laudemir C. Varanda
article en
7 citations

Abstract

In 2021, approximately 537 million people were diagnosed with diabetes mellitus. With rates expected to rise, health expenditures are projected to reach one trillion USD by 2030. Thus, measuring glucose levels is essential for rationalizing the costs of public health systems. In this context, this article presents two major contributions. First, it demonstrates that using deionized water (DI-water) as a reference for glucose sensing is not a reliable approach for representing human blood plasma (BP), as it lacks ions and suppresses essential effects such as losses. As an alternative, we investigate the use of an artificial BP solution (ABPS) that closely resembles real human BP. Characterized over a range from 500 MHz to 10 GHz, ABPS shows marginal differences in real permittivity but significant differences in imaginary permittivity compared to DI-water. The second contribution is the design of a highly sensitive microwave (MW) sensor based on double concentric circular split ring resonator (DCCSRR) on a <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$49\times 45$ </tex-math></inline-formula> mm ROGERS 5880 TM substrate. This sensor can differentiate glucose concentrations from 0 to 400 mg/dL, exceeding the relevant range for diabetic individuals (50–300 mg/dL). The DCCSRR operates at 2.48 GHz and can detect minimal concentration variations of 25 mg/dL in low concentrations, representing a significant advancement in the field. Different from most sensitive approaches available to date, this structure operates in a nonlicensed band and a fully passive form, offering flexibility for implementation and low cost. These characteristics position it as a state-of-the-art solution in MW glucose sensors.

IEEE Sensors JournalVol. 25(3)
Universidade Federal de São Carlos (BR), Institute of Mathematics and Computer Science (MD)
Fundação de Amparo à Pesquisa do Estado de São Paulo, Coordenação de Aperfeiçoamento de Pessoal de Nível Superior, Conselho Nacional de Desenvolvimento Científico e Tecnológico
Clean water and sanitation
Openalex Percentile: Top 21%
Wireless Body Area Networks
1.29
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.