Adaptive ECG Sampling – a Minimum-Error Approach

Adaptive sampling, also known as arbitrary or compressed sensing is an interesting form of concise representation of the process in a series of samples nonuniformly distributed in time.When fed with some a priori knowledge of the signal (e.g.physiological limitation) it may provide efficient usage of the data carrier, improve precise representation of selected phenomena and increased immunity to noise.In this paper we propose a smart iterative downsampling procedure to optimize sampling timepoints for best signal fidelity at given sampling rate.The method starts with identifying and sorting local extrema and then iterates the interpolation process taking alternately one half of samples as reference and minimizing the reconstruction error by adjusting timepoints' positions of the other half.The algorithm was tested with 41 records from MIT-BIH Arrhythmia Database.We found that even with 20 sps the R-peak time precision is sufficient for accurate heart rate measurement and with only 80 sps the signal reconstruction error drops below 0.1% of the full measurement scale.

Authors

Publication Details

Journal
Computing in cardiology
Published
2024-12-01
DOI
https://doi.org/10.22489/cinc.2024.391
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Adaptive ECG Sampling – a Minimum-Error Approach

Piotr Augustyniak, Debelo Oljira Hinaw
Computing in cardiology
ECG Monitoring and Analysis
article

Adaptive ECG Sampling – a Minimum-Error Approach

Piotr Augustyniak, Debelo Oljira Hinaw
article en

Abstract

Adaptive sampling, also known as arbitrary or compressed sensing is an interesting form of concise representation of the process in a series of samples nonuniformly distributed in time.When fed with some a priori knowledge of the signal (e.g.physiological limitation) it may provide efficient usage of the data carrier, improve precise representation of selected phenomena and increased immunity to noise.In this paper we propose a smart iterative downsampling procedure to optimize sampling timepoints for best signal fidelity at given sampling rate.The method starts with identifying and sorting local extrema and then iterates the interpolation process taking alternately one half of samples as reference and minimizing the reconstruction error by adjusting timepoints' positions of the other half.The algorithm was tested with 41 records from MIT-BIH Arrhythmia Database.We found that even with 20 sps the R-peak time precision is sufficient for accurate heart rate measurement and with only 80 sps the signal reconstruction error drops below 0.1% of the full measurement scale.

Computing in cardiologyVol. 51
Openalex Percentile: Top 57%
ECG Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.