Changes in Signal Morphology of Hand-held ECG Devices with Dry Electrodes

Hand-held ECG devices often comprise dry electrodes for establishing an electrical connection to the patient's body.Due to motion artifacts, static electrode charges and DC amplifier offsets, these devices usually incorporate signal conditioning, such as filter stages, which can affect the ECG signal morphology.In this work, we implemented a MATLAB Simscape-based model to simulate the electrodeskin interface of dry electrodes and the corresponding measurement pathway of a hand-held ECG measurement device to investigate the influence of skin impedance, geometric factors and electrical circuitry on the signal morphology of an Einthoven I-lead heart cycle prototype.We performed synchronized measurements with the commercially available ECG device MyDiagnostick and a Philips patient monitor MX800 for reference purposes to identify device-related signal distortions.Results indicate a significant change in signal morphology of the ECG, exhibiting an increasingly biphasic P-wave with reduced amplitude, a negative S-peak and an inverted T-wave.

Authors

Publication Details

Journal
Computing in cardiology
Published
2024-12-01
DOI
https://doi.org/10.22489/cinc.2024.434
Citations
1
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Changes in Signal Morphology of Hand-held ECG Devices with Dry Electrodes

Matthias Daniel Zink, Steffen Leonhardt, Markus Lüken
1 citations
Computing in cardiology
ECG Monitoring and Analysis
0.40
article

Changes in Signal Morphology of Hand-held ECG Devices with Dry Electrodes

Matthias Daniel Zink, Steffen Leonhardt, Markus Lüken
article en
1 citations

Abstract

Hand-held ECG devices often comprise dry electrodes for establishing an electrical connection to the patient's body.Due to motion artifacts, static electrode charges and DC amplifier offsets, these devices usually incorporate signal conditioning, such as filter stages, which can affect the ECG signal morphology.In this work, we implemented a MATLAB Simscape-based model to simulate the electrodeskin interface of dry electrodes and the corresponding measurement pathway of a hand-held ECG measurement device to investigate the influence of skin impedance, geometric factors and electrical circuitry on the signal morphology of an Einthoven I-lead heart cycle prototype.We performed synchronized measurements with the commercially available ECG device MyDiagnostick and a Philips patient monitor MX800 for reference purposes to identify device-related signal distortions.Results indicate a significant change in signal morphology of the ECG, exhibiting an increasingly biphasic P-wave with reduced amplitude, a negative S-peak and an inverted T-wave.

Computing in cardiologyVol. 51
Affordable and clean energy
Openalex Percentile: Top 29%
ECG Monitoring and Analysis
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.