Electromagnetic Design of 1.0-T Ultrashort-Cavity Intraoperative Head MRI Superconducting Magnet

This article provides an overview of the design and verification of a 1.0-T ultrashort-cavity intraoperative head magnetic resonance imaging (MRI) superconducting magnet. We proposed a linear programming method with repeated iterations for the electromagnetic design of superconducting magnets in MRI, referred to as repeated iterative linear programming (RILP). By employing RILP, we first investigated the influence of axial length on the maximum magnetic induction intensity, the number of coils, and the cost of superconducting material. The rectangularization of current clusters during the linear programming phase of the electromagnetic design was then completed. The electromagnetic force, stress, and strain were also analyzed. Finally, passive shimming within linear programming was performed on the superconducting magnet, resulting in magnetic field inhomogeneity within 20-cm diameter of spherical volume being below 8 ppm. The designed superconducting magnet meets the technical requirements for intraoperative head MRI, with a total length of 960 mm and a bore size of 650 mm.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Applied Superconductivity
Published
2024-12-16
DOI
https://doi.org/10.1109/tasc.2024.3517842
Citations
2
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.81
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Electromagnetic Design of 1.0-T Ultrashort-Cavity Intraoperative Head MRI Superconducting Magnet

Qing Zhang, Bensheng Qiu, Lei Mo, Yaen Xie et al.
2 citations
IEEE Transactions on Applied Superconductivity
Advanced MRI Techniques and Applications
0.81
article

Electromagnetic Design of 1.0-T Ultrashort-Cavity Intraoperative Head MRI Superconducting Magnet

Qing Zhang, Bensheng Qiu, Lei Mo, Yaen Xie, Xiaoyan Huang, Lin Chen, Pingping Wang, Zhengrong Liu, Huiyu Du, Yufu Zhou
article en
2 citations

Abstract

This article provides an overview of the design and verification of a 1.0-T ultrashort-cavity intraoperative head magnetic resonance imaging (MRI) superconducting magnet. We proposed a linear programming method with repeated iterations for the electromagnetic design of superconducting magnets in MRI, referred to as repeated iterative linear programming (RILP). By employing RILP, we first investigated the influence of axial length on the maximum magnetic induction intensity, the number of coils, and the cost of superconducting material. The rectangularization of current clusters during the linear programming phase of the electromagnetic design was then completed. The electromagnetic force, stress, and strain were also analyzed. Finally, passive shimming within linear programming was performed on the superconducting magnet, resulting in magnetic field inhomogeneity within 20-cm diameter of spherical volume being below 8 ppm. The designed superconducting magnet meets the technical requirements for intraoperative head MRI, with a total length of 960 mm and a bore size of 650 mm.

IEEE Transactions on Applied SuperconductivityVol. 35(2)
University of Science and Technology of China (CN), Anhui Medical University (CN), Anhui Special Equipment Inspection Institute (CN), Tianjin Hi-Tech Superconducting Electronic Technology (China) (CN), Tiandi Science & Technology (China) (CN)
Affordable and clean energy
Openalex Percentile: Top 24%
Advanced MRI Techniques and Applications
0.81
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.