Accuracy and Reliability of Grip Strength Measurements: A Comparative Device Analysis

<b>Background:</b> Grip strength is widely used as a surrogate parameter for functional status. The current gold standard, the JAMAR<sup>®</sup> Hydraulic Dynamometer (JAMAR<sup>®</sup> Hydraulic), presents challenges for individuals with painful finger joints or low grip strength. Therefore, the objective of this observational study was to assess comparability across the JAMAR<sup>®</sup> Smart, the Martin Vigorimeter and the gold standard. Additionally, the intrasubject and intersession reliability of all three devices were examined, which are essential for quality control before generating normative data. <b>Methods:</b> Forty healthy participants (aged 37.0 ± 11.3 years) were included, and a subset of 20 participants was randomly selected for retesting. Grip strength testing involved five attempts per measurement device. <b>Results:</b> Intrasubject reliability was excellent (ICC3,1: 0.91-0.97), and a strong correlation (ICC3,1: 0.90-0.98) was found between the first attempt and the best of five as well as between the best of two or three attempts and the best of five, demonstrating robust reliability across multiple measures. Intersession reliability was good to excellent (ICC3,1: 0.85-0.97) across all three devices, which was confirmed by Bland-Altman analysis. The PCC (r) revealed moderate to very strong agreement with the current gold standard JAMAR<sup>®</sup> Hydraulic (r = 0.68-0.98), with increased differences between the Martin Vigorimeter and the JAMAR<sup>®</sup> Hydraulic measurements, particularly at low and high values. <b>Conclusions:</b> The proposed devices are suitable for test-retest analysis with 2-3 attempts. Despite high correlations between all three devices, the diverging levels of agreement between the Martin Vigorimeter and the established gold standard warrant caution against using these devices interchangeably.

Authors

Institutions

Publication Details

Journal
Journal of Functional Morphology and Kinesiology
Published
2024-12-16
DOI
https://doi.org/10.3390/jfmk9040274
Citations
10
Primary Topic
Cerebral Palsy and Movement Disorders
Type
article
Field-Weighted Citation Impact
3.86
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy and Reliability of Grip Strength Measurements: A Comparative Device Analysis

Pascale Gränicher, Jaap Swanenburg, Johannes Scherr, Antoine F. Lenssen et al.
10 citations
Journal of Functional Morphology and Kinesiology
Cerebral Palsy and Movement Disorders
3.86
article

Accuracy and Reliability of Grip Strength Measurements: A Comparative Device Analysis

Pascale Gränicher, Jaap Swanenburg, Johannes Scherr, Antoine F. Lenssen, Rob A. de Bie, Jörg Spörri, Bernhard Haller, Yael Maurer
article en
10 citations

Abstract

<b>Background:</b> Grip strength is widely used as a surrogate parameter for functional status. The current gold standard, the JAMAR<sup>®</sup> Hydraulic Dynamometer (JAMAR<sup>®</sup> Hydraulic), presents challenges for individuals with painful finger joints or low grip strength. Therefore, the objective of this observational study was to assess comparability across the JAMAR<sup>®</sup> Smart, the Martin Vigorimeter and the gold standard. Additionally, the intrasubject and intersession reliability of all three devices were examined, which are essential for quality control before generating normative data. <b>Methods:</b> Forty healthy participants (aged 37.0 ± 11.3 years) were included, and a subset of 20 participants was randomly selected for retesting. Grip strength testing involved five attempts per measurement device. <b>Results:</b> Intrasubject reliability was excellent (ICC3,1: 0.91-0.97), and a strong correlation (ICC3,1: 0.90-0.98) was found between the first attempt and the best of five as well as between the best of two or three attempts and the best of five, demonstrating robust reliability across multiple measures. Intersession reliability was good to excellent (ICC3,1: 0.85-0.97) across all three devices, which was confirmed by Bland-Altman analysis. The PCC (r) revealed moderate to very strong agreement with the current gold standard JAMAR<sup>®</sup> Hydraulic (r = 0.68-0.98), with increased differences between the Martin Vigorimeter and the JAMAR<sup>®</sup> Hydraulic measurements, particularly at low and high values. <b>Conclusions:</b> The proposed devices are suitable for test-retest analysis with 2-3 attempts. Despite high correlations between all three devices, the diverging levels of agreement between the Martin Vigorimeter and the established gold standard warrant caution against using these devices interchangeably.

Journal of Functional Morphology and KinesiologyVol. 9(4)
University of Zurich (CH), TUM Klinikum (DE), Maastricht University (NL), ETH Zurich (CH), Zürich Airport (Switzerland) (CH), Technical University of Munich (DE)
Openalex Percentile: Top 5%
Cerebral Palsy and Movement Disorders
3.86
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.