Clinical Practice With Preprimed Extracorporeal Membrane Oxygenation: Safety, Sterility, and Functionality

This 5 year retrospective study presents the clinical experience with preprimed extracorporeal membrane oxygenation (ECMO) circuits used in a Single Hub Center Hospital, focusing on sterility, functionality, and safety. The ECMO program has been active since 2019, with a total of 223 circuits managed. Our preassembled and preprimed ECMO circuits were stored in a sterile environment and continuously circulated until implantation. Sterility and functionality testing was performed at the end of circuit preparation, every 7 days, and before implantation. Our results show that only 2 (0.3%) of the 570 samples tested positive for bacteria, and all implanted devices demonstrated satisfactory gas transfer performance. None of the ECMO devices demonstrated any loss of early functionality after implantation. The longest storage period of a preprimed circuit was 73 days, with no positive culture results. Our study highlights the importance of preassembled and preprimed ECMO circuits in improving clinical practice in emergency situations, highlighting their safety and potential to improve clinical practice. Furthermore, our findings suggest that standardizing guidelines for prepriming and storage of ECMO circuits can help minimize the risk of contamination.

Authors

Institutions

Publication Details

Journal
ASAIO Journal
Published
2024-12-16
DOI
https://doi.org/10.1097/mat.0000000000002361
Citations
1
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.18
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical Practice With Preprimed Extracorporeal Membrane Oxygenation: Safety, Sterility, and Functionality

Anna Papeo, Andrea Montisci, Sergio Cattaneo, Stefano Benussi et al.
1 citations
ASAIO Journal
Mechanical Circulatory Support Devices
0.18
article

Clinical Practice With Preprimed Extracorporeal Membrane Oxygenation: Safety, Sterility, and Functionality

Anna Papeo, Andrea Montisci, Sergio Cattaneo, Stefano Benussi, Carmine Puglia, Mauro Renghini, Debora Maddinelli
article en
1 citations

Abstract

This 5 year retrospective study presents the clinical experience with preprimed extracorporeal membrane oxygenation (ECMO) circuits used in a Single Hub Center Hospital, focusing on sterility, functionality, and safety. The ECMO program has been active since 2019, with a total of 223 circuits managed. Our preassembled and preprimed ECMO circuits were stored in a sterile environment and continuously circulated until implantation. Sterility and functionality testing was performed at the end of circuit preparation, every 7 days, and before implantation. Our results show that only 2 (0.3%) of the 570 samples tested positive for bacteria, and all implanted devices demonstrated satisfactory gas transfer performance. None of the ECMO devices demonstrated any loss of early functionality after implantation. The longest storage period of a preprimed circuit was 73 days, with no positive culture results. Our study highlights the importance of preassembled and preprimed ECMO circuits in improving clinical practice in emergency situations, highlighting their safety and potential to improve clinical practice. Furthermore, our findings suggest that standardizing guidelines for prepriming and storage of ECMO circuits can help minimize the risk of contamination.

ASAIO JournalVol. 71(6)
Azienda Socio Sanitaria Territoriale degli Spedali Civili di Brescia (IT)
Openalex Percentile: Top 50%
Mechanical Circulatory Support Devices
0.18
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Clinical Practice With Preprimed Extracorporeal Membrane Oxygenation: Safety, Sterility, and Functionality — Anna Papeo, Andrea Montisci, et al. · ASAIO Journal (2024) | TGRS Research Map | TGRS