Investigation of radiological properties and water equivalence of polymer gel dosimeters in the energy range 0.1–10 keV

The interaction parameters of radiation with matter are crucial for accurately determining radiation doses in diagnostic and treatment. This study examined the radiological features of normoxic polymer gels used to measure radiation doses. The mass attenuation coefficient (MAC), mass energy-absorption coefficient (MEAC), effective atomic number (Z eff ), half value layer (HVL) and mean free path (MFP) for these samples were determined in the energy range of 0.1 keV–10 keV. Z eff was established using direct and interpolation methods. The largest discrepancy for Z eff between the methods was observed for water, with a 48 % difference in the energy range of 0.3 keV–0.5 keV. MACs of the samples were compared with Monte Carlo simulation results based on FLUKA 4-3.1 code. MAC results obtained with simulation showed good agreement with Hubbell, and the maximum and minimum differences between them were seen in the FAT sample as 8.31 % and 0.08 %, respectively. Additionally, dose rate, gamma constant, and air kerma values were calculated using mass energy absorption coefficients for all samples and their water equivalence was evaluated in the relevant energy range. It was observed that the samples with radiological water equivalence and air kerma values close to 1 were PAGAT, MAGIC, MAGAS, and MAGAT.

Authors

Institutions

Publication Details

Journal
Nuclear Engineering and Technology
Published
2024-12-16
DOI
https://doi.org/10.1016/j.net.2024.103387
Citations
4
Primary Topic
Radiation Shielding Materials Analysis
Type
article
Field-Weighted Citation Impact
0.47
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigation of radiological properties and water equivalence of polymer gel dosimeters in the energy range 0.1–10 keV

Sara Parsaei, Sonika Thakur, Mehmet Büyükyıldız, Mustafa Mohammad Rafiei et al.
4 citations
Nuclear Engineering and Technology
Radiation Shielding Materials Analysis
0.47
article

Investigation of radiological properties and water equivalence of polymer gel dosimeters in the energy range 0.1–10 keV

Sara Parsaei, Sonika Thakur, Mehmet Büyükyıldız, Mustafa Mohammad Rafiei, Aytaç Levet, Parminder Kaur
article en
4 citations

Abstract

The interaction parameters of radiation with matter are crucial for accurately determining radiation doses in diagnostic and treatment. This study examined the radiological features of normoxic polymer gels used to measure radiation doses. The mass attenuation coefficient (MAC), mass energy-absorption coefficient (MEAC), effective atomic number (Z eff ), half value layer (HVL) and mean free path (MFP) for these samples were determined in the energy range of 0.1 keV–10 keV. Z eff was established using direct and interpolation methods. The largest discrepancy for Z eff between the methods was observed for water, with a 48 % difference in the energy range of 0.3 keV–0.5 keV. MACs of the samples were compared with Monte Carlo simulation results based on FLUKA 4-3.1 code. MAC results obtained with simulation showed good agreement with Hubbell, and the maximum and minimum differences between them were seen in the FAT sample as 8.31 % and 0.08 %, respectively. Additionally, dose rate, gamma constant, and air kerma values were calculated using mass energy absorption coefficients for all samples and their water equivalence was evaluated in the relevant energy range. It was observed that the samples with radiological water equivalence and air kerma values close to 1 were PAGAT, MAGIC, MAGAS, and MAGAT.

Nuclear Engineering and TechnologyVol. 57(6)
Shahid Bahonar University of Kerman (IR), Guru Nanak Dev University (IN), Bursa Technical University (TR), Atatürk University (TR)
Openalex Percentile: Top 37%
Radiation Shielding Materials Analysis
0.47
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Investigation of radiological properties and water equivalence of polymer gel dosimeters in the energy range 0.1–10 keV — Sara Parsaei, Sonika Thakur, et al. · Nuclear Engineering and Technology (2024) | TGRS Research Map | TGRS