A Comparative Analysis of Interpupillary Distance Measurement Techniques Evaluation in Modern Times: From Rulers to Apps
Measurement by different tools showed difference of IPD measurement although induced prism due to discrepancy were within the allowed tolerance of less than 0.33 prism diopter (ISO 16034:2002) for all methods. Therefore, mobile App can be efficiently used for screening purposes for many people where limited services are available. However, caution should be exercised when mobile apps are used, such as in complex and for eyes which are misaligned.
Authors
- Byoung Sun Chu (ORCID: https://orcid.org/0000-0001-9419-2194)
- Yee Rin Jung
Institutions
- Daegu Catholic University (KR)
Publication Details
- Journal
- Clinical Optometry
- Published
- 2024-12-01
- DOI
- https://doi.org/10.2147/opto.s491431
- Citations
- 4
- Primary Topic
- Advanced Optical Imaging Technologies
- Type
- article
- Field-Weighted Citation Impact
- 1.28