P056 Comparison of Breas Vivo 45 LS and ResMed Lumis ventilators in three simulated disease conditions

Abstract Background Ventilator performance has been shown to vary across conditions, ventilator type, and in common scenarios such as mask leak. Aim: This study sought to compare the performance of a Breas Medical AB (Mölnlycke, Sweden) Vivo 45 LS™ and ResMed (San Diego, CA, USA) Lumis™ ventilator. Methods Pressure, flow and volume waveforms were sampled for each ventilator at analogous settings, under three simulated clinical conditions using an ASL5000™ (Ingmar Medical, Pittsburgh, PA, USA) respiratory simulator: (A) Kyphoscoliosis (very low compliance), (B) Obesity Hypoventilation (low compliance), (C) Motor Neuron Disease (normal compliance, low muscle strength). Results Qualitative visual comparisons of representative breaths revealed similar delivered inspiratory and expiratory pressures. Flow profiles and volumes differed, with the Lumis showing shorter breath durations (range 0.1 - 0.5s) and correspondingly lower volumes compared to Vivo 45LS. Peak flows varied among the conditions, with the Vivo 45LS higher in condition A (27 vs 31 L/min) similar in condition B (45 L/min), but lower in condition C (27 vs 24 L/min). Mean inspiratory pressures were similar except in condition A, where the Vivo 45LS was lower by around 4 cmH2O. Expiratory volumes and peak pressures were comparable across both devices. Conclusions Comparative analysis of the Breas Vivo 45LS and ResMed Lumis ventilators under simulated clinical conditions highlights differences in breath duration, delivered volumes, and peak flow for analogous settings. Future research should better quantify and explore the impact of these performance variations between ventilators on clinical outcomes including ventilation efficacy, patient-ventilator synchrony, user comfort, and usage.

Authors

Institutions

Publication Details

Journal
SLEEP Advances
Published
2024-10-01
DOI
https://doi.org/10.1093/sleepadvances/zpae070.138
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

P056 Comparison of Breas Vivo 45 LS and ResMed Lumis ventilators in three simulated disease conditions

Linda Rautela, Mark E. Howard, T Churchward, Rodney Hone et al.
3 links
SLEEP Advances
Respiratory Support and Mechanisms
article

P056 Comparison of Breas Vivo 45 LS and ResMed Lumis ventilators in three simulated disease conditions

Linda Rautela, Mark E. Howard, T Churchward, Rodney Hone, Nicole Sheers, Julie Tolson, D Berlowtiz
article en

Abstract

Abstract Background Ventilator performance has been shown to vary across conditions, ventilator type, and in common scenarios such as mask leak. Aim: This study sought to compare the performance of a Breas Medical AB (Mölnlycke, Sweden) Vivo 45 LS™ and ResMed (San Diego, CA, USA) Lumis™ ventilator. Methods Pressure, flow and volume waveforms were sampled for each ventilator at analogous settings, under three simulated clinical conditions using an ASL5000™ (Ingmar Medical, Pittsburgh, PA, USA) respiratory simulator: (A) Kyphoscoliosis (very low compliance), (B) Obesity Hypoventilation (low compliance), (C) Motor Neuron Disease (normal compliance, low muscle strength). Results Qualitative visual comparisons of representative breaths revealed similar delivered inspiratory and expiratory pressures. Flow profiles and volumes differed, with the Lumis showing shorter breath durations (range 0.1 - 0.5s) and correspondingly lower volumes compared to Vivo 45LS. Peak flows varied among the conditions, with the Vivo 45LS higher in condition A (27 vs 31 L/min) similar in condition B (45 L/min), but lower in condition C (27 vs 24 L/min). Mean inspiratory pressures were similar except in condition A, where the Vivo 45LS was lower by around 4 cmH2O. Expiratory volumes and peak pressures were comparable across both devices. Conclusions Comparative analysis of the Breas Vivo 45LS and ResMed Lumis ventilators under simulated clinical conditions highlights differences in breath duration, delivered volumes, and peak flow for analogous settings. Future research should better quantify and explore the impact of these performance variations between ventilators on clinical outcomes including ventilation efficacy, patient-ventilator synchrony, user comfort, and usage.

SLEEP AdvancesVol. 5(Supplement_1)
The University of Melbourne (AU), Institute for Breathing and Sleep (AU), Austin Health (AU)
Openalex Percentile: Top 58%
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.