P034 Noxturnal cRIP: An alternative to AASM recommended flow sensors in analysis of respiratory events in polysomnography

Abstract Introduction AASM recommends the use of nasal pressure (NP), oronasal thermal flow (Th), and respiratory inductance plethysmography (RIP) for detecting and characterising respiratory events (RE) in polysomnography. A previous comparative study assessed utility of Noxturnal RIP flow (cRIPflow) as an alternative to Th for identification of RE in polysomnography. This demonstrated that cRIPflow was comparable with AASM recommendation for Th and NP (TH-NP) and indicated higher accuracy than Th in RE identification. We further investigate utility of cRIPflow between varying groups of apnoea hypopnoea index (AHI) severity. Method Respiratory scoring was performed by an experienced scorer on 10 diagnostic studies under AASM standards. Scoring was repeated using three different measurements for each study: cRIPflow only, Th only and Th-NP. AHI, central apnoea index (CAI), obstructive apnoea index (OAI), mixed apnoea index (MAI) and hypopnoea index (HI) were calculated. Data was catagorised into Group1 (AHI <30/hr) and Group2 (AHI >30/hr). Wilcoxon matched pairs test was utilised for comparison between flow measurements. Results Statistical difference found in Group1 (n=6) HI comparing cRIPflow vsTh (cRIPflow 6.08/hr+/-5.87SD, Th 4.57/hr+/-4.27SD, p=0.0312). No other statistical differences in RE identified when comparing cRIPflow vsTh, cRIPflow vs Th-NP and Th vs Th-NP within Group1 (n=6) and Group2 (n=4). Discussion This data suggests cRIPflow may be more sensitive in HI identification in lower severity AHI populations. Analysis with larger sample size would provide more insight. cRIPflow is comparative with AASM recommended flow sensors across varying AHI severity populations, suggesting an alternative method for detecting and characterising RE in polysomnography.

Authors

Institutions

Publication Details

Journal
SLEEP Advances
Published
2024-10-01
DOI
https://doi.org/10.1093/sleepadvances/zpae070.116
Primary Topic
Advanced Chemical Sensor Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

P034 Noxturnal cRIP: An alternative to AASM recommended flow sensors in analysis of respiratory events in polysomnography

Lucy Burr, Niclas Eriksson
SLEEP Advances
Advanced Chemical Sensor Technologies
article

P034 Noxturnal cRIP: An alternative to AASM recommended flow sensors in analysis of respiratory events in polysomnography

Lucy Burr, Niclas Eriksson
article en

Abstract

Abstract Introduction AASM recommends the use of nasal pressure (NP), oronasal thermal flow (Th), and respiratory inductance plethysmography (RIP) for detecting and characterising respiratory events (RE) in polysomnography. A previous comparative study assessed utility of Noxturnal RIP flow (cRIPflow) as an alternative to Th for identification of RE in polysomnography. This demonstrated that cRIPflow was comparable with AASM recommendation for Th and NP (TH-NP) and indicated higher accuracy than Th in RE identification. We further investigate utility of cRIPflow between varying groups of apnoea hypopnoea index (AHI) severity. Method Respiratory scoring was performed by an experienced scorer on 10 diagnostic studies under AASM standards. Scoring was repeated using three different measurements for each study: cRIPflow only, Th only and Th-NP. AHI, central apnoea index (CAI), obstructive apnoea index (OAI), mixed apnoea index (MAI) and hypopnoea index (HI) were calculated. Data was catagorised into Group1 (AHI <30/hr) and Group2 (AHI >30/hr). Wilcoxon matched pairs test was utilised for comparison between flow measurements. Results Statistical difference found in Group1 (n=6) HI comparing cRIPflow vsTh (cRIPflow 6.08/hr+/-5.87SD, Th 4.57/hr+/-4.27SD, p=0.0312). No other statistical differences in RE identified when comparing cRIPflow vsTh, cRIPflow vs Th-NP and Th vs Th-NP within Group1 (n=6) and Group2 (n=4). Discussion This data suggests cRIPflow may be more sensitive in HI identification in lower severity AHI populations. Analysis with larger sample size would provide more insight. cRIPflow is comparative with AASM recommended flow sensors across varying AHI severity populations, suggesting an alternative method for detecting and characterising RE in polysomnography.

SLEEP AdvancesVol. 5(Supplement_1)
Mater Health Services (AU)
Openalex Percentile: Top 79%
Advanced Chemical Sensor Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

P034 Noxturnal cRIP: An alternative to AASM recommended flow sensors in analysis of respiratory events in polysomnography — Lucy Burr, Niclas Eriksson · SLEEP Advances (2024) | TGRS Research Map | TGRS