P012 Sleeping on Clouds; a comparison of the impact of polysomnography versus remote determination on non-invasive ventilation device settings

Abstract Introduction Non-invasive ventilation (NIV) with home mechanical ventilators is an effective treatment for chronic respiratory failure. Optimal therapy requires titration of NIV settings to ensure adherence. Modern ventilators produce ventilator data for review via cloud-based platforms. The impact of relying on this data for NIV titration compared with polysomnography (PSG) is unknown. The aim of this study was to compare the agreement of ventilator settings based on summarised ventilator data and NIV PSG within participants. Methods A prospective cohort of 30 individuals with chronic respiratory failure were established on NIV. One month later, participants were administered questionnaires to assess NIV satisfaction, underwent in-lab PSG, and their data from the ventilator cloud-based platform was reviewed. Optimal NIV settings were separately determined by blinded review of the summarised ventilator data and PSG data by a respiratory and sleep physician. Results 20/30 users experienced discomfort whilst using NIV therapy. Ventilator prescription using NIV PSG and Airview data were different. Inspiratory positive airway pressure (IPAP) and pressure support (PS) were an average of 2.76cmH2O (95% confidence interval 2.0,3.5) and 1.8cmH2O (95%CI 1.1,2.5) lower based on ventilator data. Comparative analysis with Bland Altman plots demonstrated a mean difference (MD) in IPAP of 1.4cmH2O (limits of agreement (LOA) -4.47 to 7.29) and PS MD of 0.96cmH2O (LOA -3.43 to 5.36). Conclusion User reported discomfort with NIV therapy was common. Differences in IPAP and PS were observed with lower prescribed pressures when titration was performed using ventilator data compared with NIV PSG. This may influence the titration of NIV therapy; however, the clinical impact of this difference on users is currently unknown.

Authors

Institutions

Publication Details

Journal
SLEEP Advances
Published
2024-10-01
DOI
https://doi.org/10.1093/sleepadvances/zpae070.094
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

P012 Sleeping on Clouds; a comparison of the impact of polysomnography versus remote determination on non-invasive ventilation device settings

Anna Ridgers, Linda Rautela, Asha K Rajan, Warren R. Ruehland et al.
SLEEP Advances
Respiratory Support and Mechanisms
article

P012 Sleeping on Clouds; a comparison of the impact of polysomnography versus remote determination on non-invasive ventilation device settings

Anna Ridgers, Linda Rautela, Asha K Rajan, Warren R. Ruehland, Liam M. Hannan, David J. Berlowitz, M Howard, K Bruzzese
article en

Abstract

Abstract Introduction Non-invasive ventilation (NIV) with home mechanical ventilators is an effective treatment for chronic respiratory failure. Optimal therapy requires titration of NIV settings to ensure adherence. Modern ventilators produce ventilator data for review via cloud-based platforms. The impact of relying on this data for NIV titration compared with polysomnography (PSG) is unknown. The aim of this study was to compare the agreement of ventilator settings based on summarised ventilator data and NIV PSG within participants. Methods A prospective cohort of 30 individuals with chronic respiratory failure were established on NIV. One month later, participants were administered questionnaires to assess NIV satisfaction, underwent in-lab PSG, and their data from the ventilator cloud-based platform was reviewed. Optimal NIV settings were separately determined by blinded review of the summarised ventilator data and PSG data by a respiratory and sleep physician. Results 20/30 users experienced discomfort whilst using NIV therapy. Ventilator prescription using NIV PSG and Airview data were different. Inspiratory positive airway pressure (IPAP) and pressure support (PS) were an average of 2.76cmH2O (95% confidence interval 2.0,3.5) and 1.8cmH2O (95%CI 1.1,2.5) lower based on ventilator data. Comparative analysis with Bland Altman plots demonstrated a mean difference (MD) in IPAP of 1.4cmH2O (limits of agreement (LOA) -4.47 to 7.29) and PS MD of 0.96cmH2O (LOA -3.43 to 5.36). Conclusion User reported discomfort with NIV therapy was common. Differences in IPAP and PS were observed with lower prescribed pressures when titration was performed using ventilator data compared with NIV PSG. This may influence the titration of NIV therapy; however, the clinical impact of this difference on users is currently unknown.

SLEEP AdvancesVol. 5(Supplement_1)
The University of Melbourne (AU), Northern Health (AU), Institute for Breathing and Sleep (AU), Vitenparken (NO), Austin Health (AU)
Openalex Percentile: Top 58%
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.