Accuracy of NIBP Monitors Service in the Queen Sirikit Heart Center of the Northeast Hospital Thailand, According to SMM02-1 Calibration Procedure

The errors of non-invasive blood pressure (NIBP) monitors can occur in both random errors and systematic errors depending on the condition in use, length of service time, brands, models, etc. Inaccurate readings of blood pressure values can lead to wrong diagnoses and improper treatment. Therefore, a regular check of the accuracy of NIBP devices is necessary. We evaluated the accuracy of NIBP monitors in the Queen Sirikit Heart Center of the Northeast Hospital in Thailand according to SMM02-1 procedures. The test comprises static, dynamic, and pulse rate calibration, respectively. We found that long-time service monitors over 10 years tended to fail the static test by up to 25 % compared with 0–10 years monitors, which failed the test by about 14%. However, it is insignificant by the service time as 80% of monitors in this failed group are the same model. For the model study, the XA2, XB2, and Y3 models still showed high precision in both static and dynamic pressure measurement. The study showed that 60% of NIBP monitors used in Queen Sirikit Heart Center are accurate following SMM02-1. The long service time of monitors should be determined when the monitor is over 10 years. Moreover, highly accurate models of the monitors and regular calibrations are requested.

Authors

Institutions

Publication Details

Published
2024-08-08
DOI
https://doi.org/10.1109/ri2c64012.2024.10784381
Primary Topic
Cardiac Imaging and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of NIBP Monitors Service in the Queen Sirikit Heart Center of the Northeast Hospital Thailand, According to SMM02-1 Calibration Procedure

Tassanai Sanponpute, Saowalak Somjid, Theera Leeudomwon, Sayrung Moonwicha
Cardiac Imaging and Diagnostics
article

Accuracy of NIBP Monitors Service in the Queen Sirikit Heart Center of the Northeast Hospital Thailand, According to SMM02-1 Calibration Procedure

Tassanai Sanponpute, Saowalak Somjid, Theera Leeudomwon, Sayrung Moonwicha
article en

Abstract

The errors of non-invasive blood pressure (NIBP) monitors can occur in both random errors and systematic errors depending on the condition in use, length of service time, brands, models, etc. Inaccurate readings of blood pressure values can lead to wrong diagnoses and improper treatment. Therefore, a regular check of the accuracy of NIBP devices is necessary. We evaluated the accuracy of NIBP monitors in the Queen Sirikit Heart Center of the Northeast Hospital in Thailand according to SMM02-1 procedures. The test comprises static, dynamic, and pulse rate calibration, respectively. We found that long-time service monitors over 10 years tended to fail the static test by up to 25 % compared with 0–10 years monitors, which failed the test by about 14%. However, it is insignificant by the service time as 80% of monitors in this failed group are the same model. For the model study, the XA2, XB2, and Y3 models still showed high precision in both static and dynamic pressure measurement. The study showed that 60% of NIBP monitors used in Queen Sirikit Heart Center are accurate following SMM02-1. The long service time of monitors should be determined when the monitor is over 10 years. Moreover, highly accurate models of the monitors and regular calibrations are requested.

Khon Kaen University (TH), National Institute of Metrology (Thailand), King Mongkut's University of Technology North Bangkok (TH)
Good health and well-being
Openalex Percentile: Top 60%
Cardiac Imaging and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Accuracy of NIBP Monitors Service in the Queen Sirikit Heart Center of the Northeast Hospital Thailand, According to SMM02-1 Calibration Procedure — Tassanai Sanponpute, Saowalak Somjid, et al. · (2024) | TGRS Research Map | TGRS