Highly Efficient Spectral Measurement Methods Using Newly Developed High‐Throughput Magnetic Circularly Polarized Luminescence System

Magnetic circularly polarized luminescence (MCPL) spectroscopy is widely used to evaluate the luminescence dissymmetry factor (g<sub>MCPL</sub>) for compounds. However, even for the same instrument and operating conditions, the measured g<sub>MCPL</sub> is affected by errors associated with sources such as baseline drift and spectral noise, and so the range of variation of g<sub>MCPL</sub> must be considered when comparing values, which requires multiple measurements for the same sample. Also, because many samples undergo photodegradation under excitation light, it is difficult to accumulate and average spectra for samples with weak MCPL signals to improve the signal-to-noise ratio. Single measurements must therefore be performed on multiple samples and the results averaged. Furthermore, for samples with a small Stokes shift, spectral correction is required to compensate for the intensity reduction due to the inner-filter effect (IFE). Such measurements are generally performed manually and are therefore time consuming and prone to human error. Here, we demonstrate the use of a newly developed high-throughput MCPL system to automatically measure MCPL and fluorescence spectra of multiple samples of phthalocyanine complexes with high efficiency and reduced human errors. This system allows the incorporation of effective countermeasures to the issues of g<sub>MCPL</sub> variation, sample photodegradation, extremely weak MCPL signals, and the IFE.

Authors

Institutions

Publication Details

Journal
Chirality
Published
2024-12-01
DOI
https://doi.org/10.1002/chir.70001
Citations
2
Primary Topic
Luminescence and Fluorescent Materials
Type
article
Field-Weighted Citation Impact
0.23
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Highly Efficient Spectral Measurement Methods Using Newly Developed High‐Throughput Magnetic Circularly Polarized Luminescence System

Naoto Ishikawa, Taiji Oyama, Yoshitane Imai, Anas Santria et al.
2 citations
Chirality
Luminescence and Fluorescent Materials
0.23
article

Highly Efficient Spectral Measurement Methods Using Newly Developed High‐Throughput Magnetic Circularly Polarized Luminescence System

Naoto Ishikawa, Taiji Oyama, Yoshitane Imai, Anas Santria, Ken‐ichi Akao, Hiroyuki Nishikawa, Satoko Suzuki, Akio Kaneta
article en
2 citations

Abstract

Magnetic circularly polarized luminescence (MCPL) spectroscopy is widely used to evaluate the luminescence dissymmetry factor (g<sub>MCPL</sub>) for compounds. However, even for the same instrument and operating conditions, the measured g<sub>MCPL</sub> is affected by errors associated with sources such as baseline drift and spectral noise, and so the range of variation of g<sub>MCPL</sub> must be considered when comparing values, which requires multiple measurements for the same sample. Also, because many samples undergo photodegradation under excitation light, it is difficult to accumulate and average spectra for samples with weak MCPL signals to improve the signal-to-noise ratio. Single measurements must therefore be performed on multiple samples and the results averaged. Furthermore, for samples with a small Stokes shift, spectral correction is required to compensate for the intensity reduction due to the inner-filter effect (IFE). Such measurements are generally performed manually and are therefore time consuming and prone to human error. Here, we demonstrate the use of a newly developed high-throughput MCPL system to automatically measure MCPL and fluorescence spectra of multiple samples of phthalocyanine complexes with high efficiency and reduced human errors. This system allows the incorporation of effective countermeasures to the issues of g<sub>MCPL</sub> variation, sample photodegradation, extremely weak MCPL signals, and the IFE.

ChiralityVol. 36(12)
National Research and Innovation Agency (ID), Ibaraki University (JP), Kindai University (JP), The University of Osaka (JP)
Openalex Percentile: Top 49%
Luminescence and Fluorescent Materials
0.23
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.