3D-MTF of Computed Tomography (CT) Images Using a 3D-Wire Phantom: The Impact of Tube Voltage Variations

Key exposure factors, such as tube voltage are critical in CT examinations as they will impact the quality of the images produced by the CT scan. Spatial resolution is one of the parameters that can be used to determine image quality. The purpose of this study is to evaluate the impact of tube voltage in spatial resolution in three axes (x, y, z) using an in-house wire phantom. The results show that MTF 10% in the axial plane (x-y) is approximately identical (approximately 0,70 mm) for various tube voltages 80, 100, 120, and 140 kV. The same results were observed at the coronal and sagittal plane (z-axis), although they exhibit more variation than axial plane the differences in values between tube voltages are not significant.

Authors

Institutions

Publication Details

Journal
International Journal of Scientific Research in Science and Technology
Published
2024-12-10
DOI
https://doi.org/10.32628/ijsrst24114311
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

3D-MTF of Computed Tomography (CT) Images Using a 3D-Wire Phantom: The Impact of Tube Voltage Variations

Choirul Anam, Wahyu Setia Budi, Betha Sri Wulandari, Yuliana Lakapu
International Journal of Scientific Research in Science and Technology
Advanced X-ray and CT Imaging
article

3D-MTF of Computed Tomography (CT) Images Using a 3D-Wire Phantom: The Impact of Tube Voltage Variations

Choirul Anam, Wahyu Setia Budi, Betha Sri Wulandari, Yuliana Lakapu
article en

Abstract

Key exposure factors, such as tube voltage are critical in CT examinations as they will impact the quality of the images produced by the CT scan. Spatial resolution is one of the parameters that can be used to determine image quality. The purpose of this study is to evaluate the impact of tube voltage in spatial resolution in three axes (x, y, z) using an in-house wire phantom. The results show that MTF 10% in the axial plane (x-y) is approximately identical (approximately 0,70 mm) for various tube voltages 80, 100, 120, and 140 kV. The same results were observed at the coronal and sagittal plane (z-axis), although they exhibit more variation than axial plane the differences in values between tube voltages are not significant.

International Journal of Scientific Research in Science and TechnologyVol. 11(6)
Diponegoro University (ID)
Openalex Percentile: Top 79%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

3D-MTF of Computed Tomography (CT) Images Using a 3D-Wire Phantom: The Impact of Tube Voltage Variations — Choirul Anam, Wahyu Setia Budi, et al. · International Journal of Scientific Research in Science and Technology (2024) | TGRS Research Map | TGRS