Chromatographic Retention Index for the Identification of Illicit Substances

New Psychoactive Substances have a high introduction rate into the market and represent a huge challenge from a regulatory and forensic point of view. Thus, the need for a method capable of identifying NPS, without the availability of certified analytical standards, arises. The methodology of the work relies on the development of a GC/MS method; its validation; and the calculation of the Kovats’ Retention Index. This index was selected because it was reported as a tool for high-precision identification. A total of twenty-two substances were used: AM-2201, MAM-2201, JWH-081, JWH-210, N-ethylpentylone, 5-MAPB, 2-FA, 25C-NBOMe, o-CPP, p-CPP, U-47700, 5-MeO-MiPT, 5-IAI, Salvia divinorium, methiopropamine, THC, heroin, amphetamine, methamphetamine, MDMA, cocaine, and ephedrine. The method presented one Selectivity limitation (coelution of methiopropamine and methamphetamine). Resolution values were higher than 1.25, indicating signal separation. Separation Factor, Number of Theoretical Plates, and Tailing provided satisfactory results. Evaluation of Intermediate Precision indicated Relative Standard Deviations (RSD) that varied from 0.02 to 0.29%, lower than the literature. Robustness evaluation using a Fractional Factorial Design identified that changing the column polarity was the most influential factor, while column brand, gas flow, split rate, injector temperature, and ramp temperature did not interfere significantly. Retention Index was calculated without overlap and the Global Confidence Interval varied from 0,5 to 20,6 Retention Index Unity and the RSD varied from 0,02 to 0,29%, inferior to the range found in the literature. The results of the development and validation indicate the adequacy of the method that uses Kovats’ Retention Index to identify NPS. The collection of Index calculated can be updated and made available to forensic institutes to be used as a tool in NPS identification.

Authors

Institutions

Publication Details

Journal
Emerging Trends in Drugs Addictions and Health
Published
2024-12-01
DOI
https://doi.org/10.1016/j.etdah.2023.100092
Primary Topic
Analytical Methods in Pharmaceuticals
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Chromatographic Retention Index for the Identification of Illicit Substances

Jorge Ambrósio, Jorge J. Zacca, Élvio D. Botelho, Marcelo Pereira de Souza et al.
Emerging Trends in Drugs Addictions and Health
Analytical Methods in Pharmaceuticals
article

Chromatographic Retention Index for the Identification of Illicit Substances

Jorge Ambrósio, Jorge J. Zacca, Élvio D. Botelho, Marcelo Pereira de Souza, Gabriella H. Giudice, Angelo H. L. Machado, Luíza N.B. Caldas
article en

Abstract

New Psychoactive Substances have a high introduction rate into the market and represent a huge challenge from a regulatory and forensic point of view. Thus, the need for a method capable of identifying NPS, without the availability of certified analytical standards, arises. The methodology of the work relies on the development of a GC/MS method; its validation; and the calculation of the Kovats’ Retention Index. This index was selected because it was reported as a tool for high-precision identification. A total of twenty-two substances were used: AM-2201, MAM-2201, JWH-081, JWH-210, N-ethylpentylone, 5-MAPB, 2-FA, 25C-NBOMe, o-CPP, p-CPP, U-47700, 5-MeO-MiPT, 5-IAI, Salvia divinorium, methiopropamine, THC, heroin, amphetamine, methamphetamine, MDMA, cocaine, and ephedrine. The method presented one Selectivity limitation (coelution of methiopropamine and methamphetamine). Resolution values were higher than 1.25, indicating signal separation. Separation Factor, Number of Theoretical Plates, and Tailing provided satisfactory results. Evaluation of Intermediate Precision indicated Relative Standard Deviations (RSD) that varied from 0.02 to 0.29%, lower than the literature. Robustness evaluation using a Fractional Factorial Design identified that changing the column polarity was the most influential factor, while column brand, gas flow, split rate, injector temperature, and ramp temperature did not interfere significantly. Retention Index was calculated without overlap and the Global Confidence Interval varied from 0,5 to 20,6 Retention Index Unity and the RSD varied from 0,02 to 0,29%, inferior to the range found in the literature. The results of the development and validation indicate the adequacy of the method that uses Kovats’ Retention Index to identify NPS. The collection of Index calculated can be updated and made available to forensic institutes to be used as a tool in NPS identification.

Emerging Trends in Drugs Addictions and HealthVol. 4
National Water Agency (BR)
Openalex Percentile: Top 78%
Analytical Methods in Pharmaceuticals
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.