Accurate Intramyocardial Hemorrhage Assessment with Fast, Free-running, Cardiac Quantitative Susceptibility Mapping

Purpose To evaluate the performance of a high-dynamic-range quantitative susceptibility mapping (HDR-QSM) cardiac MRI technique to detect intramyocardial hemorrhage (IMH) and quantify iron content using phantom and canine models. Materials and Methods A free-running whole-heart HDR-QSM technique for IMH assessment was developed and evaluated in calibrated iron phantoms and 14 IMH female canine models. IMH detection and iron content quantification performance of this technique was compared with the conventional iron imaging approaches, R2*(1/T2*) maps, using measurements from ex vivo imaging as the reference standard. Results Phantom studies confirmed HDR-QSM's accurate iron content quantification and artifact mitigation ability by revealing a strong linear relationship between iron concentration and QSM values (<i>R</i><sup>2</sup>, 0.98). In in vivo studies, HDR-QSM showed significantly improved image quality and susceptibility homogeneity in nonaffected myocardium by alleviating motion and off-resonance artifacts (HDR-QSM vs R2*: coefficient of variation, 0.31 ± 0.16 [SD] vs 0.73 ± 0.36 [<i>P</i> < .001]; image quality score [five-point Likert scale:], 3.58 ± 0.75 vs 2.87 ± 0.51 [<i>P</i> < .001]). Comparison between in vivo susceptibility maps and ex vivo measurements showed higher performance of HDR-QSM compared with R2* mapping for IMH detection (area under the receiver operating characteristic curve, 0.96 vs 0.75; <i>P</i> < .001) and iron content quantification (<i>R</i><sup>2</sup>, 0.71 vs 0.14). Conclusion In a canine model of IMH, the fast and free-running cardiac QSM technique accurately detected IMH and quantified intramyocardial iron content of the entire heart within 5 minutes without requiring breath holding. <b>Keywords:</b> High-Dynamic-Range Quantitative Susceptibility Mapping, Myocardial Infarction, Intramyocardial Hemorrhage, MRI <i>Supplemental material is available for this article.</i> ©RSNA, 2024.

Authors

Institutions

Publication Details

Journal
Radiology Cardiothoracic Imaging
Published
2024-12-01
DOI
https://doi.org/10.1148/ryct.230376
Citations
4
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.62

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accurate Intramyocardial Hemorrhage Assessment with Fast, Free-running, Cardiac Quantitative Susceptibility Mapping

Hao Ho, Hsin‐Yao Lin, Ghazal Yoosefian, Hsu‐Lei Lee et al.
4 citations
Radiology Cardiothoracic Imaging
Advanced MRI Techniques and Applications
1.62
article

Accurate Intramyocardial Hemorrhage Assessment with Fast, Free-running, Cardiac Quantitative Susceptibility Mapping

Hao Ho, Hsin‐Yao Lin, Ghazal Yoosefian, Hsu‐Lei Lee, Shing Fai Chan, Xingmin Guan, Hui Han, Xiaoming Bi, Gregory Anthony, Dhirendra Singh, Yuheng Huang, Keyur Vora, Xinheng Zhang, Fei Han, Behzad Sharif, Rohan Dharmakumar, Khalid Youssef, Li-Ting Huang, Hsin‐Jung Yang, Debiao Li, Anthony Christodoulou
article en
4 citations

Abstract

Purpose To evaluate the performance of a high-dynamic-range quantitative susceptibility mapping (HDR-QSM) cardiac MRI technique to detect intramyocardial hemorrhage (IMH) and quantify iron content using phantom and canine models. Materials and Methods A free-running whole-heart HDR-QSM technique for IMH assessment was developed and evaluated in calibrated iron phantoms and 14 IMH female canine models. IMH detection and iron content quantification performance of this technique was compared with the conventional iron imaging approaches, R2*(1/T2*) maps, using measurements from ex vivo imaging as the reference standard. Results Phantom studies confirmed HDR-QSM's accurate iron content quantification and artifact mitigation ability by revealing a strong linear relationship between iron concentration and QSM values (<i>R</i><sup>2</sup>, 0.98). In in vivo studies, HDR-QSM showed significantly improved image quality and susceptibility homogeneity in nonaffected myocardium by alleviating motion and off-resonance artifacts (HDR-QSM vs R2*: coefficient of variation, 0.31 ± 0.16 [SD] vs 0.73 ± 0.36 [<i>P</i> < .001]; image quality score [five-point Likert scale:], 3.58 ± 0.75 vs 2.87 ± 0.51 [<i>P</i> < .001]). Comparison between in vivo susceptibility maps and ex vivo measurements showed higher performance of HDR-QSM compared with R2* mapping for IMH detection (area under the receiver operating characteristic curve, 0.96 vs 0.75; <i>P</i> < .001) and iron content quantification (<i>R</i><sup>2</sup>, 0.71 vs 0.14). Conclusion In a canine model of IMH, the fast and free-running cardiac QSM technique accurately detected IMH and quantified intramyocardial iron content of the entire heart within 5 minutes without requiring breath holding. <b>Keywords:</b> High-Dynamic-Range Quantitative Susceptibility Mapping, Myocardial Infarction, Intramyocardial Hemorrhage, MRI <i>Supplemental material is available for this article.</i> ©RSNA, 2024.

Radiology Cardiothoracic ImagingVol. 6(6)
Cedars-Sinai Medical Center (US), University of California, Los Angeles (US), Mackay Memorial Hospital (TW), Institute of Statistical Science, Academia Sinica (TW), National Cheng Kung University Hospital (TW), Indiana University School of Medicine, Indiana University – Purdue University Indianapolis (US)
National Institutes of Health
Good health and well-being
Openalex Percentile: Top 15%
Advanced MRI Techniques and Applications
1.62
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.